Research Article

Realization of the Radiance Scale Using Transfer Function of the Laser-Based Optical System

Table 4

k = 2 expanded standard uncertainty budget for 632.8 nm in the 11 mrad FOV.

Uncertainty component (Xi)SymMeasured values (xi)UnitStandard measurement uncertainty UnitSensitivity coefficient (Ci)UnitUncertainty contribution Variance

Trap detector responsivityR3.3170e – 01A/W3.6487e − 03A/W−1.0316e + 00W2/A·m2·sr·nm−3.7640e − 031.4167e − 05
Geometric constantΔC6.1421e – 07m2·sr1.2536e − 08m2·sr−5.5711e + 05W/(m2·sr2·nm−6.9841e − 034.8777e − 05
Detector output signalIs2.3811e – 07A7.8377e − 11A1.4371e + 06W/A·m2·sr·nm1.1263e − 041.2687e − 08
Spectrometer output signalIm2.8013e − 10A1.9801e − 13A−1.2215e + 09W/A·m2·sr·nm−2.4187e − 045.8500e − 08
Unknown output signalIU1.7669e − 10A6.4885e − 13A1.9366e + 09W/A·m2·sr·nm1.2566e − 031.5790e − 06
Wavelength band passΔλ2.6500e + 00nm1.0000e − 02nm−1.2912e − 01W/m2·sr·nm2−1.2912e − 031.6673e − 06
Correction factorsδR1.2301e + 003.7718e − 041.0000e + 003.7718e − 041.4227e − 07
RepeatabilityδR3.6793e − 01W/m2·sr·nm5.5423e − 04W/m2·sr·nm1.0000e + 005.5423e − 043.0717e − 07
Sum (W/m2·sr·nm)6.6404e − 05
Total measurement uncertainty (W/m2·sr·nm) 0.0081
Expanded measurement uncertainty (W/m2·sr·nm) 0.0163