Realization of the Radiance Scale Using Transfer Function of the Laser-Based Optical System
Table 5
k = 2 expanded standard uncertainty for 11 mrad FOV in terms of both GUM and GUM Supplement-1.
| Wavelength (nm) | GUM | Monte Carlo | Ratio of the two calculations | Calculated radiance (W/m2·sr·nm) | k = 2 expanded measurement uncertainty (W/m2·sr·nm) | Calculated radiance (W/m2·sr·nm) | k = 2 expanded measurement uncertainty (W/m2·sr·nm) |
| 457 | 0.11952 | 0.0058 | 0.11953 | 0.0057 | 1.0229 | 477 | 0.14892 | 0.0072 | 0.14887 | 0.0071 | 1.0198 | 488 | 0.16825 | 0.0119 | 0.16819 | 0.0108 | 1.0998 | 514 | 0.19524 | 0.0094 | 0.19483 | 0.0093 | 1.0086 | 543 | 0.24632 | 0.0122 | 0.24708 | 0.0121 | 1.0099 | 594 | 0.32095 | 0.0153 | 0.32041 | 0.0153 | 1.0033 | 604 | 0.33359 | 0.0165 | 0.33469 | 0.0162 | 1.0204 | 612 | 0.32635 | 0.0155 | 0.32543 | 0.0155 | 0.9980 | 633 | 0.34347 | 0.0163 | 0.34237 | 0.0163 | 0.9993 | 1064 | 0.25589 | 0.0150 | 0.25424 | 0.0146 | 1.0253 |
|
|