Summary of electroacoustic verification results for all HI cases (1 through 5). Summaries are shown across stimuli (2 kHz and 4 kHz tone bursts) and hearing aid conditions (NLFC enabled and no-NLFC). Results are compared to the corresponding cortical auditory evoked potentials (CAEP) judged as present or absent per condition.

Case 2 kHz NLFC2 kHz no-NLFC 4 kHz NLFC 4 kHz no-NLFC
SL (dB)CAEPSL (dB)CAEPSL (dB)CAEPSL (dB)CAEP

130.29Present30.24Present11.47Present5.36Present
219.36Present19.86Present10.06Present0.39Absent
313.1Present13.36Present16.29Present6.19Absent
49.68Present9.66Present−7.24Present−18.25Absent
50.02Present6.51Present−10Present−18.6Absent