The Reliability of Pharyngeal High Resolution Manometry with Impedance for Derivation of Measures of Swallowing Function in Healthy Volunteers

Table 1

Intrarater reliability of 1st analysis versus 2nd analysis for the Primary and Repeat Studies. Data are average ICC [range ICC].

ICC for 1st analysis versus 2nd analysis
Primary StudyRepeat Study

Contractility
 PhPP0.98 [0.90–1.00]0.93 [0.67–1.00]
 PhCI0.97 [0.91–1.00]0.87 [0.51–0.98]
 hPP0.94 [0.84–0.99]0.91 [0.71–0.98]
 UBP1.00 [1.00-1.00]1.00 [0.99–1.00]
 UPP0.93 [0.75–1.00]0.98 [0.89–1.00]
 UCI0.88 [0.71–0.96]0.99 [0.97–1.00]
Intrabolus pressure
 PhIBP1.00 [1.00–1.00]0.97 [0.88–1.00]
 hIBP0.98 [0.95–1.00]0.99 [0.97–0.99]
 UIRP0.99 [0.97–1.00]1.00 [0.99–1.00]
Flow timing
 PhDCL0.96 [0.79–1.00]0.97 [0.91–1.00]
 hDCL0.85 [0.29–0.99] 0.93 [0.83–0.99]
 hFI0.99 [0.99–1.00]0.99 [0.96–1.00]
Global function
 Swallow Risk Index0.99 [0.94–1.00]0.94 [0.80–1.00]