Research Article
Comparison of CMM and Micro-CT Volumetric Analysis of Polyethylene Tibial Knee Inserts in Total Knee Replacement
Table 3
CMM and Micro-CT measurement versus gravimetric measurement.
| CMM measurement versus gravimetric measurement |
| ā | CCC | Precision | Accuracy | Statistic | 0.990 | 1.000 | 0.991 | 95% CL | 0.979 | 0.999 | 0.980 |
| Micro-CT measurement versus gravimetric measurement | ā | CCC | Precision | Accuracy |
| Statistic | 0.975 | 0.999 | 0.976 | 95% CL | 0.947 | 0.996 | 0.949 |
|
|