Optical Path Length Calibration: A Standard Approach for Use in Absorption Cell-Based IR-Spectrometric Gas Analysis

Figure 6

OPC data as a function of processed to retrieve the optical path lengths of the CUTs A (a) and B (b) of = (21.89 ± 0.08) m and = (0.819 ± 0.0029) m, respectively. For (a) a mixture of about 20 mmol/mol CO2 in nitrogen was used, whereas for (b) a mixture of 10 mmol/mol CH4 in nitrogen was probed for the OPC.
132607.fig.006a
(a) 21 m gas cell
132607.fig.006b
(b) 82 cm gas cell

We are committed to sharing findings related to COVID-19 as quickly and safely as possible. Any author submitting a COVID-19 paper should notify us at help@hindawi.com to ensure their research is fast-tracked and made available on a preprint server as soon as possible. We will be providing unlimited waivers of publication charges for accepted articles related to COVID-19. Sign up here as a reviewer to help fast-track new submissions.