Optical Path Length Calibration: A Standard Approach for Use in Absorption Cell-Based IR-Spectrometric Gas Analysis

Table 3

Uncertainty budget of for gas cell B (82 cm) derived by means of (2) with CH4 being the probed sensor molecule. The index column is the relative contribution of the properties variance to the variance of [11].

Input quantityValueRelative uncertainty/%Index/%

p 244 hPa0.160.0
T 294.8 K0.08 0.1
0.00402 cm−10.40.6
0.01 mol/mol0.993.7
8.289 × 10−22
cm−1/(cm−2·molec)
595.4
0.805 m5