Redefining Onyx HD 500 in the Flow Diversion Era

Academic Editor: Karthikeshwar Kasirajan
Received08 Jun 2011
Revised24 Aug 2011
Accepted24 Aug 2011
Published27 Oct 2011

Abstract

We report the largest US case series results using Onyx HD-500 (EV3), a new liquid embolic agent, in the successful treatment of 21 patients with wide-neck intracranial aneurysms (mean size 4.5 mm), which are at increased risk of incomplete occlusion or recanalization with standard endovascular intervention utilizing detachable platinum coils. All aneurysms were located in the anterior circulation, and three aneurysms presented as acute subarachnoid hemorrhages. Complete aneurysm occlusion was present in 19 of 21 patients (90%). On six-month followup, one patient with an initially small residual neck progressed to total occlusion. Aneurysm recanalization was not detected in any patients on mean follow up of 8.9 months in 11 patients. Four patients experienced transient neurologic deficits in the immediate postoperative period and one in a delayed fashion. Embolization with the liquid embolic agent Onyx appears to be a safe and effective endovascular modality of treatment for wide-neck aneurysms or recurrent aneurysms that had previously failed treatment with detachable coils.

1. Introduction

Treatment modalities achieving endosaccular occlusion of aneurysms continue to rapidly evolve. As technologies progress it is imperative to demonstrate improved efficiency, patient care, and clinical outcomes. Liquid embolic agents designed for aneurysm treatment potentially carry additional therapeutic benefits over coil embolization including improved volumetric aneurysm filling and endothelialization across the aneurysm neck which can lead to reduced recanalization rates with fewer procedures and lower expenses. The use of liquid embolic agents stands in contrast to the recent strategy of flow diversion for the treatment of intracranial aneurysms.

The Onyx liquid embolic system (EV3, Inc. Irvine, Calif, USA) is a combination agent of ethylene-vinyl alcohol copolymer, dimethylsulfoxide, and tantalum that has been primarily used in the endovascular treatment of arteriovenous malformations. Since 1999, Onyx has also been used in cerebral aneurysms and has been studied in several European and South American case series [13]. Since then, the technique for its use and types of polymers used have been refined with a newer high-viscosity agent, Onyx HD 500, designed specifically for the endovascular treatment of intracranial aneurysms. We report the largest published series of intracranial aneurysms treated in the United States with Onyx HD 500 for aneurysms with a wide neck (>4 mm or dome-to-neck ration <2).

2. Methods

This study was a US single center prospective registry study that included 21 consecutive patients with intracranial aneurysms treated by embolization with Onyx HD 500 between March 2008 and April 2011 in a single institution. Similar to patients enrolled in the CAMEO and other Onyx trials, patients were selected based on aneurysms characteristics that were difficult to treat by conventional methods of coil embolization with a neck wider than 4 mm or dome-to-neck ratios less than 2 [3]. Several aneurysms were also selected that had recurred following prior coil embolization. Data was collected on patients prior to the procedure, at the time of the procedure, at discharge, and again at 6–12-month followup. Clinical data that were evaluated included preoperative comorbidities and presenting symptoms, modified Rankin scale, cranial nerve deficits, any adverse effects, and changes in neurologic status as seen in Table 1.


Case no.Procedure dateAgeSexPresenting symptomsHx of SAHPre-Op mRSAneurysm locationAneurysm sideType of aneurysmAneurysm dome height (mm)Aneurysm dome width (mm)Aneurysm neck length (mm)Parent artery diameter (mm)

103/18/0845F IncidentalNone0Ophthalmic arteryLeftSidewall4.44.15.54.5
203/18/0854F IncidentalPrevious SAH0Posterior communicatingRightSidewall3.62.62.44.0
305/13/0883FAneurysm rupture/SAHSAH5Posterior communicatingRightSidewall10.012.17.54.4
410/10/0854FAneurysm rupture/SAHSAH5Superior hypophysealRightSidewall3.44.34.45.7
511/17/0935FHeadacheNone0OphthalmicLeftSidewall6.05.04.04.0
612/09/0944FIncidentalPrevious SAH0Posterior communicatingRightSidewall3.03.03.04.8
712/21/0961FOcculomotor nerve palsyNone0Cavernous carotidRightSidewall20.015.08.03.8
812/23/0935FHeadacheNone0Superior hypophysealRightSidewall4.03.02.54.5
912/24/0959MOcculomotor nerve palsyNone1Cavernous carotidLeftSidewall27.018.07.05.0
1001/26/1050FIncidentalPrevious SAH0Posterior communicatingLeftSidewall3.02.22.04.0
1101/29/1057FVisual field deficitNone0OphthalmicLeftSidewall6.06.54.54.0
1204/28/1054FIncidentalPrevious SAH0Posterior communicating ArteryRightSidewall2.83.02.03.9
1309/24/1085FVisual field deficitNone1OphthalmicLeftSidewall12.014.04.54.3
1410/07/1068FHeadacheNone0Clinoid segmentRightSidewall8.012.06.64.0
1510/20/1060FOcculomotor nerve palsyNone0Cavernous CarotidLeftSidewall8.09.03.02.0
1611/01/1068FAneurysm rupture/SAHNone5Cavernous carotidRightDissecting5.04.05.04.0
1711/19/1052FHeadacheNone0OphthalmicLeftSidewall5.38.09.25.0
1802/28/1166FHeadacheNone0Superior hypophysealLeftSidewall5.64.53.04.5
1904/04/1153FHeadacheAVM1OpthalmicLeftSidewall5.15.63.34.5
2004/18/1169FHeadacheNone0Superior hypophysealRightSidewall3.93.33.95.0
2104/21/1153MIncidentalNone0Anterior choriodalLeftSidewall5.74.53.54.0

All unruptured aneurysms were pretreated with clopidogrel 75 mg per day and aspirin 81 mg for at least 10 days before the procedure. Patients with subarachnoid hemorrhage (SAH) were loaded with 600 mg of clopidogrel during the procedure through the nasogastric tube. In only one patient was Enterprise stent (Codman Neurovascular, Raynham, MA, USA) assistance utilized at the time of onyx embolization; however, five patients were coiled previously, and one of them had a stent-assisted coiling that presented on follow up with a recurrence. In no patient was adjunctive coil embolization employed. The procedure time from microcatheter insertion to removal ranged from 25 to 120 minutes with a mean of 56 minutes. The average time under fluoroscopy was 75 minutes. The total volume of onyx used ranged from 0.2 mL to 8.4 mL with a mean of 1.2 mL.

All patients were treated under general anesthesia and full systemic anticoagulation with heparin. Since we perform all of our endovascular aneurysm treatments under general anesthesia, neurophysiological monitoring was used as a substitute for the neurological exam during these procedures. Electrophysiologic monitoring was performed for continuous BAERs, SSEPs, and EEG. Activated clotting times were maintained greater than 250 seconds throughout the entire case. Our technique used for Onyx injection is similar to those previously described as seen in Figures 1 and 2 [28]. After the Onyx embolic agent is dry heated and shaken, a 6F guide catheter is placed in the internal carotid artery and a DMSO-compatible Hyperglide balloon (EV3, Irvine, Calif, USA) positioned and inflated to allow its middle portion to cover the neck of the aneurysm. The microcatheters utilized in this study to access the aneurysm were the DMSO-compatible Rebar 14 (EV3) or Echelon 10, which are placed within the mid-portion of the aneurysm. Next, with the balloon inflated, contrast agent is slowly injected to ensure an effective seal test, demonstrating stasis of the contrast in the aneurysm and allowing estimation of the amount of balloon inflation needed for complete aneurysm occlusion while avoiding leakage of the embolic agent. The balloon is deflated at this point.

Next, saline is flushed through the microcatheter, and the dead space is filled with DMSO. Onyx is then carefully injected until the material approaches the end of the microcatheter at the recommended rate of 0.2 mL/2 min. With the DMSO then displaced, the balloon is inflated after 0.15 mL of onyx is injected to fill part of the microcatheter dead space. At this point, Onyx is injected to fill the aneurysm sac with the balloon inflated. In our institution, with the use of intraoperative neurophysiologic monitoring, the balloon is continuously inflated as long as the monitoring is stable. We do not interrupt our injection to deflate the balloon. Onyx is allowed to laminate around the balloon, creating an alpha sign, in the parent artery to cover the aneurysm neck, ensuring a complete and durable occlusion and reducing the risk of aneurysm recurrence. With confirmation of adequate aneurysm obliteration, there is a 10-minute hold period to allow complete solidification of the cast with the balloon deflated. The balloon is then inflated for counter resistance and the microcatheter removed by gentle but brisk traction.

3. Results

Patient’s ages ranged from 35–85 years with a mean age of 57 as seen in Table 1. Three patients presented as acute Hunt and Hess Grade IV Subarachnoid Hemorrhage, 4 had a previous history of SAH, and the remaining 11 aneurysms were unruptured. Five patients presented with cranial nerve deficits. Of these 3 had cranial nerve III palsies from cavernous carotid aneurysms, and two had visual field deficits from ophthalmic artery aneurysms. Eleven patients presented with headaches, while 3 patients’ aneurysms were completely incidental findings on MRI/A. Five patients had been previously treated with coil embolization but had recanalized, and 1 patient was previously coiled with stent assistance incompletely. All patients underwent cerebral digital subtraction angiography prior to treatment.

All of the aneurysms were located in the anterior circulation arising on a segment of the internal carotid artery. One aneurysm was located in the clinoid segment, 4 superior hypophyseal, 4 cavernous carotid, 5 posterior communicating, 6 ophthalmic, and 1 in the anterior choroidal segment. Ten aneurysms were located off the left ICA, while 9 were right sided. Fourteen aneurysms were small (<8 mm), six were large (8–24), and one was giant (≥24 mm). Eleven aneurysms had necks greater than 4 mm with a mean neck length of 4.5 mm. The average parent artery diameter was 4.3 mm.

All 21 patients were successfully treated with Onyx HD 500, and 19 of 21 had complete occlusion as the immediate angiographic result with two patients displaying a small residual neck as seen in Table 2. One Grade IV Subarachnoid hemorrhage patient required decompressive hemicraniectomy postoperatively due to increased intracranial pressures, and another patient with a grade V SAH ultimately progressed to brain death. Three patients had minor unintentional embolic migration into the parent artery without flow limitation. On followup parent artery stenosis was not observed. Initially, 5 of the 21 patients experienced transient neurologic adverse events as described in Table 3. All patients undergoing their six-month followup had improved to their baseline neurologic condition. Ten of 21 patients (47%) had followup angiography within one year (one patient with follow-up Magnetic Resonance Angiography) and all of these patients achieved complete occlusion.


Case no.Plavix 650 mg givenBalloon usedMicrocatheterAccess catheterTotal volume (mL) Onyx injected (incl dead space)Immediate angiographic Raymond scaleImmediate occlusionAdverse eventsmRS at discharge

1Yes HyperglideRebar 148F Guiding catheter0.30Complete occlusion100%No adverse events0
2Yes HyperglideRebar 148F Guiding catheter0.30Complete occlusion100%No adverse events0
3No HyperglideRebar 148F Guiding catheter0.50Complete occlusion100%Required decompressive hemicraniectomy5
4Yes HyperglideRebar 148F Guiding catheter0.20Complete occlusion100%Death6
5Yes HyperglideRebar 148F Guiding catheter0.70Complete occlusion100%No adverse events0
6Yes HyperglideRebar 148F Guiding catheter0.60Small residual neck95–99%Femoral access complication0
7Yes HyperglideRebar 148F Guiding catheter4.80Complete occlusion100%Worsening of oculomotor deficit0
8Yes HyperglideRebar 148F Guiding catheter0.80Complete occlusion100%TIAs0
9Yes HyperglideRebar 148F Guiding catheter8.40Complete occlusion100%Worsening of oculomotor deficit1
10Yes HyperglideRebar 148F Guiding catheter0.50Complete occlusion100%No adverse events0
11Yes HyperglideRebar 148F Guiding catheter0.85Small residual neck95–99%TIAs1
12Yes HyperglideRebar 148F Guiding catheter0.43Complete occlusion100%Transient visual disturbance1
13Yes HyperglideRebar 148F Guiding catheter2.70Complete occlusion100%No adverse events1
14Yes HyperglideEchelon 108F Guiding catheter1.40Complete occlusion100%Asymptomatic Cervical ICA dissection0
15Yes HyperglideEchelon 148F Guiding catheter1.20Complete occlusion100%No adverse events0
16No HyperglideEchelon 106F Guiding catheter0.55Complete occlusion100%No adverse events5
17Yes HyperglideEchelon 106F Guiding catheter0.35Complete occlusion100%No adverse events0
18Yes HyperglideEchelon 106F Guiding catheter0.30Complete occlusion100%No adverse events0
19Yes HyperglideRebar 106F Guiding catheter0.40Complete occlusion100%No adverse events0
20Yes HyperglideEchelon 106F Guiding catheter0.30Complete occlusion100%No adverse events0
21Yes HyperglideEchelon 148F Guiding catheter0.40Complete occlusion100%No adverse events0


Case no.Timing of followupDegree of change of followup angiogramFollowup raymond scaleFollowup angio % occlusionFollowup mRS

19 monthsNo changeComplete occlusion100%0
27 monthsNo changeComplete occlusion100%0
612 monthsNo changeComplete occlusion100%0
79 monthsNo changeComplete occlusion100%0
811 monthsNo changeComplete occlusion100%0
96 monthsNo changeComplete occlusion100%1
108 monthsNo changeComplete occlusion100%0
119 months Increased occlusionComplete occlusion100%0
1212 monthsNo changeComplete occlusion100%0
146 months (MRA)No changeComplete occlusion100%0

4. Discussion

Since the publication of the results of ISAT and ISUIA, the development of endovascular techniques utilizing platinum coils has been a dominant tool in the armamentarium of neurovascular interventionalists to achieve complete occlusion and prevent aneurysm rupture [9, 10]. However, aneurysms with wide necks (>4 mm), dome-to-neck ratios less than 1.5, or very large in dome height that have been treated with coils are associated with high rates of recanalization. This deficiency in the endovascular treatment of aneurysms has led to the development of newer techniques and materials to completely occlude these difficult-to-treat cerebral aneurysms.

Onyx liquid embolic is a biocompatible nonadhesive agent composed of ethylene-vinyl alcohol copolymer and dimethyl sulfoxide (DMSO) solvent added to tantalum powder to make the solution radioopaque. Being a liquid embolic agent allows Onyx to fill 100% of the aneurysm sac volume. Originally, different polymer concentrations Onyx 18 and 34 (6% and 8%) were used in AVM embolization. An Onyx 12% polymer concentration was then initially used for aneurysm embolization until Onyx HD (high density; 500 cP) was developed with a 20% copolymer concentration with an increased viscosity to lessen the likelihood of reflux into the parent artery. Onyx utilizes a balloon remodeling technique that may enable better control of the neck and parent vessel compared to traditional coiling [7]. There are also reports of Onyx inducing more neoendothelialization over the neck of the aneurysm relative to coils [6, 11].

There have been previous reports of the endovascular use of Onyx in intracranial aneurysms beginning with Mawad et al. describing embolization of 11 large and giant intracavernous aneurysms with an initial 82% complete occlusion and 91% at followup in 6 or 12 months [5]. They did not report any recanalization. The European CAMEO trial was the first multicenter trial with Onyx embolization that described 100 treated aneurysms with an occlusion rate at 1-year followup ranging from 93% in small aneurysms to 57% in giant aneurysms with an overall occlusion rate of 79% [3]. They retreated 10% of patients and reported an 8% permanent morbidity and 2% mortality rate. Following these trials with promising initial results, Onyx HD 500 was developed with a greater viscosity to reduce the risk of parent vessel occlusion and distal migration that were seen in these prior trials. Weber et al. was the first European trial to report their initial use Onyx HD 500 in 22 aneurysms with an immediate occlusion rate of 82% and 91% at approximately one year [8]. Temporary or minor complications were reported in 14% of patients, more permanently in 4.5% and 11% of aneurysms recanalized. They did not report any deaths. The largest South American Onyx HD 500 trial to date was published by Piske et al. which reported on 84 wide-neck aneurysms with a 66% total occlusion on short-term follow up, 85% at 6 months, and 90% at 18 months [6]. They reported progression from incomplete to complete occlusion in 68%, with recanalization in 5% in which 2 patients (3%) required retreatment. Their reported mortality was 2.9% with a permanent morbidity of 7%. Simon et al. was the first US group to discuss their preliminary experience with Onyx HD 500 in 12 intracranial aneurysms [7]. They described initial angiographic results in which all aneurysms were 90% or more occluded. All patients with followup six-month DSA displayed durable occlusion. They experienced one complication with ophthalmic occlusion and no mortalities.

Our results in 21 patients with wide-neck aneurysms treated with Onyx HD 500 are compared favorably with those of the CAMEO trial and other series and better than those reported for coil embolization as seen in Table 4. 19 patients achieved immediate complete occlusions, and one initially near-complete occlusion progressed to complete occlusion. Eight patients (38%) experienced adverse events following the procedure including worsening of preexisting cranial neuropathies in 2 patients, transient visual disturbance in 1 patient, TIAs in 1 patient, and femoral access complications in 1 patient; 2 patients with SAH had complications following the procedure that were due to their primary disease (1 grade IV patient required hemicraniectomy and 1 grade V patient died). Of the initial adverse events in unruptured aneurysms, none produced permanent morbidity with all neurological sequelae resolving completely at discharge or on followup.


AuthorsDate publishedJournal# Aneurysms treatedPrimary outcome% Mortality% Permanent disability attributed to procedure

Mawad et al. [5]Mar, 2002J Neurosurg11GOS9%9%
Molyneux et al. [3]Jan, 2004AJNR97Percent occlusion2%8%
Lubicz et al. [2]Apr, 2005AJNR41mGOS3%7%
Weber et al. [8]Sept, 2005AJNR22Percent occlusion0%0%
Cekirge et al. [1]Jan, 2006Neuroradiology100mRS3%8%
Piske et al. [6]May, 2009Neurosurgery84Percent occlusion3%7%
Simon et al. [7]Sept, 2010Neurosurgery12Durable occlusion0%8%

These results are encouraging and must be weighed against recent aneurysm treatment strategies utilizing endoluminal treatments and flow diversion. These techniques aim to induce aneurysmal thrombosis by diverting the flow away from the aneurysm; however, this obliteration effect takes place in a delayed fashion and requires weeks to months for full effect. In the setting of acute SAH, flow diversion is relatively contraindicated when considering that the risk of rerupture is close to 20% in 2 weeks and 50% by 6 months. Also, flow diversion stents can not be used at arterial bifurcations. Further, with early reports on the Pipeline stent describing some delayed aneurysm ruptures and intra parenchymal hemorrhages flow diversion stent’s efficacy and adverse effect profile cannot be assumed until larger series are reported [12].

At this point in time, Onyx HD 500 has still some specific indications, such as multilobulated aneurysms close to a bifurcation (like our case in Figure 3), wide-neck aneurysms below 10 mm (considering the FDA has only approved the pipeline embolization device for aneurysms >10 mm), the setting of subarachnoid hemorrhage, in previously stented aneurysms where pipeline cannot provide the diversion desired, and in small recurrences after coil embolization where Onyx HD 500 can effectively seal the neck of the aneurysm.

In our series, we treated three SAH patients (2 grade IV and one grade V) and immediately achieved complete aneurysm occlusion in all patients. Two patients were eventually discharged to inpatient rehabilitation facilities with permanent disabilities, and the other ultimately proceeded to brain death. Similar to flow diversion techniques however, Onyx embolization requires antiplatelet agents preoperatively and postoperatively, making its use in patients with an acute aneurysm rupture less than ideal. However, the morbidity and mortality of an otherwise unprotected and recently ruptured aneurysm and the lack of other options at the time (based upon aneurysm characteristics) made the use of Onyx HD 500 a necessity.

Taken together, our overall results using Onyx HD 500 were encouraging, but will necessarily need to be confirmed with larger series and in comparison to newer self-expanding stent-assisted coiling and flow diversion techniques.

5. Conclusion

Onyx HD 500 appears to be a safe and effective endovascular modality of treatment for wide-neck aneurysms that are not amenable for, or had previously failed treatment with detachable coils. With an adequate patient selection, this method may be a better alternative than coiling in regards to packing density, intraoperative rupture, and recurrence rate. In the flow diversion era, Onyx HD 500’s use needs to be redefined; however it remains a valuable tool in the endovascular armamentarium.

References

  1. H. S. Cekirge, I. Saatci, S. Geyik, K. Yavuz, H. Öztürk, and G. Pamuk, “Intrasaccular combination of metallic coils and Onyx liquid embolic agent for the endovascular treatment of cerebral aneurysms,” Journal of Neurosurgery, vol. 105, no. 5, pp. 706–712, 2006. View at: Google Scholar
  2. B. Lubicz, M. Piotin, C. Mounayer, L. Spelle, and J. Moret, “Selective endovascular treatment of intracranial aneurysms with a liquid embolic: a single-center experience in 39 patients with 41 aneurysms,” American Journal of Neuroradiology, vol. 26, no. 4, pp. 885–893, 2005. View at: Google Scholar
  3. A. J. Molyneux, S. Cekirge, I. Saatci, and G. Gál, “Cerebral aneurysm multicenter European Onyx (CAMEO) trial: results of a prospective observational study in 20 European centers,” American Journal of Neuroradiology, vol. 25, no. 1, pp. 39–51, 2004. View at: Google Scholar
  4. H. S. Cekirge, I. Saatci, M. H. Ozturk et al., “Late angiographic and clinical follow-up results of 100 consecutive aneurysms treated with Onyx reconstruction: largest single-center experience,” Neuroradiology, vol. 48, no. 2, pp. 113–126, 2006. View at: Publisher Site | Google Scholar
  5. M. E. Mawad, S. Cekirge, E. Ciceri, and I. Saatci, “Endovascular treatment of giant and large intracranial aneurysms by using a combination of stent placement and liquid polymer injection,” Journal of Neurosurgery, vol. 96, no. 3, pp. 474–482, 2002. View at: Google Scholar
  6. R. L. Piske, L. H. Kanashiro, E. Paschoal, C. Agner, S. S. Lima, and P. H. Aguiar, “Evaluation of onyx HD-500 embolic system in the treatment of 84 wide-neck intracranial aneurysms,” Neurosurgery, vol. 64, no. 5, pp. E865–E875, 2009. View at: Publisher Site | Google Scholar
  7. S. D. Simon, E. Eskioglu, A. Reig, and R. A. Mericle, “Endovascular treatment of side wall aneurysms using a liquid embolic agent: a US single-center prospective trial,” Neurosurgery, vol. 67, no. 3, pp. 855–860, 2010. View at: Publisher Site | Google Scholar
  8. W. Weber, R. Siekmann, B. Kis, and D. Kuehne, “Treatment and follow-up of 22 unruptured wide-necked intracranial aneurysms of the internal carotid artery with Onyx HD 500,” American Journal of Neuroradiology, vol. 26, no. 8, pp. 1909–1915, 2005. View at: Google Scholar
  9. A. Molyneux, R. Kerr, I. Stratton et al., “International subarachnoid aneurysm trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised trial,” The Lancet, vol. 360, no. 9342, pp. 1267–1274, 2002. View at: Publisher Site | Google Scholar
  10. D. O. Wiebers, J. P. Whisnant, J. Huston III et al., “Unruptured intracranial aneurysms: natural history, clinical outcome, and risks of surgical and endovascular treatment,” The Lancet, vol. 362, no. 9378, pp. 103–110, 2003. View at: Publisher Site | Google Scholar
  11. Y. Murayama, F. Vinuela, S. Tateshima, F. Vinuela Jr., and Y. Akiba, “Endovascular treatment of experimental aneurysms by use of a combination of liquid embolic agents and protective devices,” American Journal of Neuroradiology, vol. 21, no. 9, pp. 1726–1735, 2000. View at: Google Scholar
  12. I. Szikora, Z. Berentei, Z. Kulcsar et al., “Treatment of intracranial aneurysms by functional reconstruction of the parent artery: the Budapest experience with the pipeline embolization device,” American Journal of Neuroradiology, vol. 31, no. 6, pp. 1139–1147, 2010. View at: Publisher Site | Google Scholar

Copyright © 2012 Richard Tyler Dalyai et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.


More related articles

1537 Views | 910 Downloads | 5 Citations
 PDF  Download Citation  Citation
 Download other formatsMore
 Order printed copiesOrder

Related articles

We are committed to sharing findings related to COVID-19 as quickly and safely as possible. Any author submitting a COVID-19 paper should notify us at help@hindawi.com to ensure their research is fast-tracked and made available on a preprint server as soon as possible. We will be providing unlimited waivers of publication charges for accepted articles related to COVID-19. Sign up here as a reviewer to help fast-track new submissions.