Research Article

Use of the AFX Stent Graft in Patients with Extremely Narrow Aortic Bifurcation: A Multicenter Retrospective Study

Table 3

Key procedural data. Data is presented as with the min–max range for abdominal aortic aneurysm (AAA) and penetrating aortic ulcer (PAU). Student’s -test or Mann–Whitney test was applied according to the Kolmogorov-Smirnov normality test to identify differences between AAA and PAU patients ().

TargetAAA ()PAU ()Ψ valueTotal ()
Min–max rangeMin–max rangeMin–max range

Procedure time (min)75-22354-17154-223
Fluoroscopy time (min)9.5-40.03.8–35.33.8-40
Contrast agent (ml)13-12015-7013-120

ΨIncluding one patient who was treated for a type Ib endoleak (EL) using the AFX stent graft (initial treatment for PAU).