International Federation of Clinical Chemistry (IFCC): Scientific Division, Committee on pH, Blood Gases and Electrolytes: Guidelines for Transcutaneouspo2 and pco2 Measurement

Department of Clinical Chemistry, Herning Hospital, Herning DK 7400, Denmark

Copyright © 1989 Hindawi Publishing Corporation. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

This document provides guidelines for the terminology, methodology, and for the interpretation of data obtained from the use of skin (transcutaneous) po2 and pco2 electrodes. The transcutaneous technique has found special application in newborn infants. The causes of analytical bias with respect to arterial blood gas values, and imprecision obtained with transcutaneous pco2 electrodes, are reviewed. Electrode temperatures above 44°C should not be used routinely, and, at a measuring temperature of 44°C, the measuring site should be changed at least every 4 h to avoid skin burns.