Refinement and evaluation of an automated mass spectrometer for nitrogen isotope analysis by the Rittenberg technique

Department of Agronomy, University of Illinois, Turner Hall, 1102 South Goodwin Avenue, Urbana, Illinois 61801, USA

Copyright © 1991 Hindawi Publishing Corporation. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

An apparatus designed to automatically perform hypobromite oxidations of ammonium salt samples for nitrogen isotope analyses with a mass spectrometer was modified to improve performance and reduce analysis time. As modified, reference N2 is admitted to the mass spectrometer between samples from a dedicated inlet manifold, for calibration at the same pressure as that of the preceding sample. Analyses can be performed on samples containing 10 μg to 1 mg of N (or more), at a rate of up to 350 samples/day. When operated with a double-collector mass spectrometer, the standard deviation at the natural abundance level (10 analyses, 50-150 μg N) was <0.0001 atom % 15N. Very little memory was observed when natural abundance samples (0.366 atom % 15N) were analysed. following samples containing 40 atom % 15N. Analyses in the range, 0.2 to 1 atom % 15N (50-150 μg N), were in good agreement with manual Rittenberg analyses (1 mg N) using a dual-inlet system, and precision was comparable. For enrichments of 2 to 20 atom % 15N, automated analyses were slightly lower than manual analyses, which was attributed to outgassing of N2 from the plastic microplate used to contain samples.