Research Article

A Study of the Critical Uncertainty Contributions in the Analysis of PCBs in Ambient Air

Table 1

GC-MS method and optimisation parameters.

Gas chromatograph parameters

Injection modeSplitless
Injection port temperature275°C
Injection volume2 L
GC to MS transfer line temperature290°C
Carrier gas and flow rateHelium; 1.8 mL⋅min−1
ColumnHP-5MS [(5% phenyl-) methyl-polysiloxane],
30 m × 0.25 mm × 0.25 m (film thickness)
Temperature program(1) 70°C-hold for 2 minutes
(2) ramp to 150°C at 15°C⋅min−1-no hold
(3) ramp to 300°C at 10°C⋅min−1-hold for 5 minutes
[Total time of program = 27.3 minutes]

Mass spectrometer parameters

Ion source temperature250°C
Quadrupole temperature150°C
Electron multiplier voltage2800 V
Other tuning parametersOptimised automatically