A Study of the Critical Uncertainty Contributions in the Analysis of PCBs in Ambient Air
Table 1
GC-MS method and optimisation parameters.
| Gas chromatograph parameters |
| Injection mode | Splitless | Injection port temperature | 275°C | Injection volume | 2 L | GC to MS transfer line temperature | 290°C | Carrier gas and flow rate | Helium; 1.8 mL⋅min−1 | Column | HP-5MS [(5% phenyl-) methyl-polysiloxane], | 30 m × 0.25 mm × 0.25 m (film thickness) | Temperature program | (1) 70°C-hold for 2 minutes | (2) ramp to 150°C at 15°C⋅min−1-no hold | (3) ramp to 300°C at 10°C⋅min−1-hold for 5 minutes | [Total time of program = 27.3 minutes] |
| Mass spectrometer parameters |
| Ion source temperature | 250°C | Quadrupole temperature | 150°C | Electron multiplier voltage | 2800 V | Other tuning parameters | Optimised automatically |
|
|