Liang Wei, Xinshen Zhang, Yichun Dai, Jin Huang, Yong-hong Xie, Kai Xiao, "Online Preconcentration and Determination of Trace Amounts of Zinc in Nature Waters", Journal of Analytical Methods in Chemistry, vol. 2008, Article ID 742319, 5 pages, 2008. https://doi.org/10.1155/2008/742319

Online Preconcentration and Determination of Trace Amounts of Zinc in Nature Waters

Academic Editor: Peter Stockwell
Received08 Oct 2007
Accepted02 Jan 2008
Published19 Mar 2008

Abstract

A simple, sensitive, reliable and flexible flow injection spectrophotometric method is proposed for on-line preconcentration and determination of trace amounts of zinc in water. At the presence of Tween-80 in pH 9.3 buffer solutions, the shade of color of Zn (II)-PAN complex is in a linear relation to the zinc amount at the point of the maximum absorption peak of 560 nm. The optimal experimental conditions, including reaction conditions and preconcentration conditions, had been obtained. The linear range of the proposed method was between 2.0 and 360  𝜇 g L 1 and the detection limit was 0.42  𝜇 g L 1 . The relative standard deviation was 3.55% and 2.14% for 5.0  𝜇 g L 1 and 50  𝜇 g L 1 of zinc standard solution ( 𝑛 = 8 ). The method had been successfully applied to zinc determination in water samples and the analytical results were satisfactory.

1. Introduction

Zinc is an essential element in the nutrition of animals, human beings, and plants [1, 2]. Zinc plays an important role in replications, gene expressions, and the metabolism of nucleic acids and different proteins. However, zinc can be toxic when exposures exceed physiological needs. Zinc is present in food, soil, and water. Hence, determination of zinc (II) is indispensable. There are many methods for zinc determination such as spectrophotometry [35], atomic absorption spectrometry [6, 7], flame atomic absorption spectroscopy [8], graphite furnace atomic absorption spectrometry [9, 10], inductively coupled plasma atomic emission spectroscopy [11], and inductively coupled plasma mass spectrometry [12]. The above methods mentioned are of high sensitivity and accuracy. But some are expensive, some are time-consuming, some are troublesome, and others are toxic. Modern environmental monitor calls for automatic, safe, and simple analytical methods.

Preconcentration technique [1316] can improve the detection limit as well as the selectivity of the method. Flow injection analysis (FIA) [1719] can improve analytical procedures such as high analytical throughput, precision, accuracy, low reagents, and sample consumption. Therefore, a new method was proposed for on-line preconcentration and spectrophotometric determination of low-concentration Zn (II) in water. The proposed method is based on the fact that Zn reacts with PAN to produce Zn (II)-PAN complex formation in the pH 9.3 buffer solution. To avoid the extraction procedure, it has been found that the PAN reacts with zinc to form a water-soluble complex at the presence of Tween-80.

2. Experiment

2.1. Apparatus

A model ZJ-la automatic metallic element analyzer developed and produced successfully by Professor Xinshen Zhang of our own laboratory was used. The analyzer has the functions of flow injection analysis (FIA), ion chromatography, and automatic sample injection. The function of automatic reference FIA was applied to determine Zn (II). Two analytical pumps (Shanghai Huxi Analytical Instrument Plant, Shanghai, China) of FIA were used to deliver solutions. One was used to deliver elution solution and reagent solution while the other delivered sample solution. Polytetrafluoroethylene (PTFE) tubing of 0.5 mm in internal diameter was used as the channels for all solutions to circulate.

Ultraviolet and visible spectra were obtained using a Spectrumlab S54 (Lengguang Technology Co., Ltd, Shanghai, China). The absorbance intensity was recorded at 560 nm on an IBM-PC. Data acquisition and processing were performed with HW-2000 chromatography software (Qianpu software Co., Ltd, Shanghai, China) running under Windows XP.

2.2. Reagents

All reagents used including 1-(2-pyridineazo)-2-naphthol (PAN) (Tianjin Ruijinte Chemical Reagent Plant, Tianjin, China), Tween-80, sodium tetraborate, o-phenanthroline, DDTC, and sodium hexametaphosphate (Chengdu Fang ZhouChemical Reagent Plant, Chengdu, China), oxalic acid dihydrate, lithium hydroxide monohydrate, ammonium thiocyanate, and citric acid (Beijing DonghuaChemical Reagent Plant, Beijing, China) were of analytical grade and all solutions were prepared with deionized water.

2.3. Preparation of Solution

Zinc stock solution: zinc stock standard solutions at a concentration of 1000 mg L−1 were obtained from China Environmental Monitoring Terminus, Beijing, China. Working solutions were prepared by suitable dilution of the stock solution.

Mixed chromogenic reagent solution (R): 10 mL of 4 × 1 0 3  mol L−1 PAN ethanol solution, 20 mL of 8% Tween-80, 20 mL of pH 9.3 of sodium tetraborate suffer solution, 1.5 mL of 10% sodium hexametaphosphate, 1 mL of 0.1% DDTC, and 2 mL of o-phenanthroline were mixed and the mixture solution was diluted to 100 mL.

Elution solution: a total of 1.26 g of oxalic acid dihydrate, 1.576 g of citric acid monohydrate, and 1.15 g of lithium hydroxide monohydrate were dissolved in 1000 mL of water.

2.4. Preparation of Preconcentration Column

The preconcentration column developed by our laboratory was filled with macroporous spherical resin with 4-(2-pyridylazo) resorcinol (PAR) functional group. The PAR functional group can form chelate complex with determined metal ion in the preconcentration column and the chelate complex can be eluted by the eluent solution of oxalic acid-citric acid-lithium hydroxide. The filling granularity of resin was 80–100  𝜇 m. The preconcentration column length was 40 mm and its internal diameter was 5 mm.

By filling the preconcentration column, mix resin and deionized water into pasty matter and drop slowly into column with burette. After sedimentating for a moment, resin was dropped slowly until the preconcentration column became full. Then, cover strainer, cover on the column, and wash a few minutes with deionized water for further work.

2.5. Analytical Procedure

The FIA manifold used was outlined in Figure 1 with PTFE pipes. The sample solution was injected into preconcentration column to preconcentrate with a six-way injection valve. When the instrument was put into analyzing position, the concentrated Zn (II) was eluted by eluent solution, pushed to the three-way cock, and mixed with chromogenic reagent solution in the reaction coil. The Zn (II)-PAN complex was then formed and carried into the flow cell. The absorbance intensity of Zn (II)-PAN complex was determined at 560 nm by detector and transformed to a signal of the peak which was recorded by HW-2000 software on a PC.

3. Results and Discussion

3.1. Effect of Reaction Conditions
3.1.1. Effect of pH

In the flow injection analysis, the absorbance of complex is affected by pH of the reaction medium. The effect of pH on the Zn (II)-PAN complex was tested in the pH range 4–11. The test results showed that the maximum peak height could be obtained when pH range was between 8 and 10. Thus, pH 9.3 of borate solution was selected as the buffer solution.

3.1.2. Effect of PAN Concentration

The reagent concentration was another important factor in flow injection analysis. The effect of PAN concentration on the peak height were tested from 5 × 1 0 5  mol L−1 to 1 . 2 × 1 0 3  mol L−1 in Figure 2. The results showed that the peak height rose when concentration of PAN changed from 5 × 1 0 5  mol L−1 to 4 × 1 0 4  mol L−1, and the peak height declined when concentration of PAN changed from 4 × 1 0 4  mol L−1 to 1 . 2 × 1 0 3  mol L−1. Therefore, 4 × 1 0 4  mol L−1 PAN was selected for further work.

According to the reaction kinetics, the reaction rate will inevitably increase when increasing concentrations of the reagent. Thus, the absorbency will increase when increasing concentrations of PAN. However, the reaction rate will not increase when concentrations of the reagent reaches a certain value. Therefore, when PAN consumption reaches a certain value, the concentration of PAN which reacts with zinc should be certain because zinc concentration is fixed in the sample. Thereby, the absorbency will not increase when increasing PAN concentration. On the contrary, the background color (baseline noises) will increase because the free state PAN in the system will increase.

3.1.3. Effect of Tween-80 Concentration

The Zn (II)-PAN complex was hydrophobic. Thus, Tween-80 was selected as reagent which could increase solubility of the system. The test results in Figure 3 showed that the peak height increased when concentration of Tween-80 changed from 1.0% to 1.6% and that the peak height did not increase even when concentration of Tween-80 changed from 1.6% to 2.2%. Therefore, 1.6% Tween-80 was selected as the optimal concentration for the further work.

3.1.4. Effect of Flow Rate

In the FIA, the flow rate could affect the sensitivity of the proposed method. In the study, the flow rate was adjusted by the interior diameter of pump pipe. Therefore, the effect of the flow rate of reagent solution on the peak height was studied from 0.10 mL min−1 to 0.72 mL min−1. The test results showed that the best flow rate could be obtained when the flow rate of reagent (R) was set at 0.23 mL min−1.

3.1.5. Effect of the Reaction Coil Length

In the flow injection analysis, the reaction time is adjusted by the length of reaction coil. If reaction coil was too short, the reagents would react incompletely. If reaction coil was too long, it would be time-consuming. Therefore, reaction coil length from 1 to 5 m was tested. The study results in Figure 4 showed that 3 m of the reaction coil length was the optimum and further increase of reaction coil length did not increase peak height. Thus, 3 m of the reaction coil length was chosen as optimal parameter.

3.2. Effect of Preconcentration Conditions
3.2.1. Effect of Filling Granularity and Length of Preconcentration Column

Length of preconcentration column and filling granularity could affect the effect of preconcentration. In the study, filling granularity of 8 0 1 0 0 𝜇 m was selected and length of concentration column from 20 mm to 100 mm was tested. The results showed that the shorter the length of concentration column, the higher the sensitivity. However, if the length of concentration column was too short, the concentration of preconcentration Zn (II) in samples was not adequate, which lowered the sensitivity. The highest sensitivity was obtained when the length of concentration column was 40 mm. Thus, 40 mm of concentration column length was selected for further study.

3.2.2. Effect of Preconcentration Time

Preconcentration time can affect analysis sensitivity and sampling frequency. Preconcentration time from 2 minutes to 13 minutes was tested. The results in Figure 5 showed that the preconcentration time was longer, the zinc concentration of the preconcentration column would be higher which causes the analysis sensitivity to be higher but the sampling frequency to be lower. Taken a comprehensive consideration of analysis sensitivity and sampling frequency, preconcentration time of 5 minutes was decided for further work.

3.2.3. Effect of Preconcentration Rate

Either too high or too low preconcentration rate was not proper. Too high-preconcentration rate would increase analysis sensitivity but augment pressure of preconcentration column which would cause leakage or pump pipes rupture. Too low-preconcentration rate would reduce analysis sensitivity and sampling frequency but augment the preconcentration time. Taken a comprehensive consideration of analysis sensitivity and sampling frequency, preconcentration rate of 1.5 min L−1 was chosen for further work.

3.2.4. Effect of Eluent Concentration

If eluent concentration was too low, Zn (II) in preconcentration column would elute incompletely. If eluent concentration was too high, it would waste reagent. In the study, nitric acid and oxalic acid-citric acid-lithium hydroxide were tested as eluents. The experimental results showed the sensitivity of the eluent of oxalic acid-citric acid-lithium hydroxide was higher than that of nitric acid. Thus, oxalic acid-citric acid-lithium hydroxide was selected as eluent solution and its concentrations were 0.01 mol L−1, 0.0075 mol L−1 and 0.05 mol L−1, respectively.

3.2.5. Effect of Eluent Flow Rate

The sensitivity was greatly affected by flow rate of eluent. If flow rate was too high, the sensitivity augmented and baseline noises increased. If flow rate was too slow, Zn (II) in concentration column would elute incompletely, which caused sensitivity decline. The result showed that flow rate of 0.5 mL min−1 of eluent was optimal for further work.

3.3. Effects of Foreign Ions

The interference of some possibly coexisting foreign metallic ions was examined to make sure the proposed method can be applied to determine Zn (II) in water. The tolerance limit was defined as the concentration of added solutions causing less than ±5% relative error during the determination of 10 μg L−1 of Zn (II). No interference was observed from a great deal of K+, Na+, NH4+, Mg2+, Ca2+, Cr3+, Cr6+, Al3+, Co2+, Hg2+, As3+, and so forth. However, the metallic ions of Ni2+, Cd2+, Mn2+, Pb2+, Cu2+, and Fe2+ interfered. The interference was eliminated (100 μg L−1 of each ion of Ni2+, Cd2+, Mn2+, Pb2+, Cu2+, and Fe2+) by adding 1.5 mL of 10% sodium hexametaphosphate, 1 mL of 0.1% DDTC, and 2 mL of o-phenanthroline as appropriate masking reagents into mixed chromogenic reagent solution. Thus, the results showed that the proposed method had good selectivity.

3.4. Linearity, Precision, and Detection Limit of the Proposed Method

To assess the reproducibility and accuracy of the proposed method, under the selected conditions above, tests of linearity, precision, and detection limit were conducted. The linearity was tested by a series of working standard solutions of Zn (II) ranging from 0.5  𝜇 g L−1 to 400  𝜇 g L−1. Test showed that the peak height versus Zn (II) concentration was linear within the range of 2.0 g L−1 to 360  𝜇 g L−1. The curve was H = 0.427C + 3.638 (H: peak height; C: concentration of Zn (II) ( 𝜇 g L−1)) and the correlation coefficient obtained was 0.9995 and the detection limit (3 multiples of baseline noises) was 0.42  𝜇 g L−1. The test of precision was conducted by eight injections of 5.0  𝜇 g L−1 and 50  𝜇 g L−1 of Zn (II) standard solution, respectively. The results showed that the relative standard deviation (RSD) was calculated as 3.55% and 2.14%, respectively.

3.5. Application

The proposed method was applied to determine Zn (II) in water samples using the ZJ-la automatic metallic element analyzer. The analysis results of samples were shown in Table 1.


SampleMeasuredSpikedFoundRecovery
( 𝜇 g L−1)( 𝜇 g L−1)( 𝜇 g L−1)(%)

River water 1#38.5030.0067.6397.1
River water 2#59.2650.00108.5698.6
Lake water4.485.009.72104.8
Spring water7.2610.0017.1298.6

To prove the accuracy of the method, a test of recovery would be conducted by accurately adding known concentration of Zn (II) into sample whose original Zn (II) had been determined. The test results in Table 1 showed that the recovery was between 97.1% and 104.8% and analytical results of the proposed method were satisfactory. Therefore, the proposed method was suitable for determination of trace amounts of Zn (II) in water.

4. Conclusion

The proposed method was significant with regard to the development of a simple, reliable, and sensitive flow injection method for on-line preconcentration and determination of trace amounts of zinc (II). High sensitivity, selectivity, broad determination range (2.0  𝜇 g L−1 to 360  𝜇 g L−1) of Zn (II), low-detection limit (0.42  𝜇 g L−1), and fully automated analysis were just some of the advantages of the proposed method. Especially, compared with other existing methods of zinc preconcentration, eluent concentration is even low to mmol L−1. The detection limit of the proposed method can be further improved by increasing preconcentration time. The model ZJ-la automatic metallic elements analyzer employed in this study was cheap, small, and easy to move. The proposed method has been successfully applied to determine zinc (II) in water. The method is suitable for determination of Zn (II) in both water samples and other samples for its good selectivity and the above mentioned advantages.

Acknowledgment

The authors are very grateful for the sponsorship of the Hi-Tech Research and Development Program of China (863 Program) (Project no. 2005AA639290).

References

  1. L. A. Brunborg, K. Julshamn, R. Nortvedt, and L. Frøyland, “Nutritional composition of blubber and meat of hooded seal (Cystophora cristata) and harp seal (Phagophilus groenlandicus) from Greenland,” Food Chemistry, vol. 96, no. 4, pp. 524–531, 2006. View at: Publisher Site | Google Scholar
  2. R. J. Van Saun, “Nutritional diseases of South American camelids,” Small Ruminant Research, vol. 61, no. 2-3, pp. 153–164, 2006. View at: Publisher Site | Google Scholar
  3. K. J. Reddy, J. R. Kumar, C. Ramachandraiah, T. Thriveni, and A. V. Reddy, “Spectrophotometric determination of zinc in foods using N-ethyl-3-carbazolecarboxaldehyde-3-thiosemicarbazone: evaluation of a new analytical reagent,” Food Chemistry, vol. 101, no. 2, pp. 585–591, 2007. View at: Publisher Site | Google Scholar
  4. G. Gumus, H. Filik, and B. Demirata, “Determination of bismuth and zinc in pharmaceuticals by first derivative UV-visible spectrophotometry,” Analytica Chimica Acta, vol. 547, no. 1, pp. 138–143, 2005. View at: Publisher Site | Google Scholar
  5. H. Sedaira, “Simultaneous determination of manganese and zinc in mixtures using first- and second-derivative spectrophotometry,” Talanta, vol. 51, no. 1, pp. 39–48, 2000. View at: Publisher Site | Google Scholar
  6. M. A. Taher, “Atomic absorption spectrometric determination of ultra trace amounts of zinc after pre-concentration with the ion pair of 2-(5-bromo-2-pyridylazo)-5-diethylaminophenol and ammonium tetraphenylborate on microcrystalline naphthalene or by column method,” Talanta, vol. 52, no. 2, pp. 181–188, 2000. View at: Publisher Site | Google Scholar
  7. S.-Y. Chen, C.-N. Chang, C.-L. Li, and S.-J. J. Tsai, “Binary modifiers for the determination of zinc in pure copper and nickel-based alloy by longitudinal Zeeman electrothermal atomic absorption spectrometry,” Analytica Chimica Acta, vol. 550, no. 1-2, pp. 156–163, 2005. View at: Publisher Site | Google Scholar
  8. J. Chen and K. C. Teo, “Determination of cadmium copper lead and zinc in water samples by flame atomic absorption spectrometry after cloud point extraction,” Analytica Chimica Acta, vol. 450, no. 1-2, pp. 215–222, 2001. View at: Publisher Site | Google Scholar
  9. H. Shangda and S. Kunyauh, “Direct determination of zinc in seawater by graphite furnace atomic absorption spectrometry,” Spectrochimi Acta, vol. 50B, pp. 837–846, 1995. View at: Google Scholar
  10. H. Minamisawa, K. Murashima, M. Minamisawa, N. Arai, and T. Okutani, “Determination of indium by graphite furnace atomic absorption spectrometry after coprecipitation with chitosan,” Analytical Sciences, vol. 19, no. 3, pp. 401–404, 2003. View at: Publisher Site | Google Scholar
  11. A. D. Eaton, L. S. Clescer, and A. E. Greenberg, Standard Methods for the Examination of Water and Wastewater, American Public Health Association, Washington, DC, USA, 19th edition, 1995.
  12. J. Fang, J. P. Chen, Y. H. Shu, and H. P. Wang, “ICP-MS determination of As Cd Cu Pb and Zn in chrysanthenum tea,” Cinese Journal of Physical Testing and Chemical Analysis B, no. 12, pp. 22–26, 2005. View at: Google Scholar
  13. V. A. Lemos, W. N. L. dos Santos, J. S. Santos, and M. B. de Carvalho, “On-line pre-concentration system using a minicolumn of polyurethane foam loaded with Me-BTABr for zinc determination by flame atomic absorption spectrometry,” Analytica Chimica Acta, vol. 481, no. 2, pp. 283–290, 2003. View at: Publisher Site | Google Scholar
  14. S. K. Mehta, A. K. Malik, B. Singh, and A. L. J. Rao, “Development of new adsorbent chitin for column pre-concentration and spectrophotometric trace determination of Ziram and Zineb in synthetic, commercial samples and food-stuffs,” Talanta, vol. 67, no. 4, pp. 725–729, 2005. View at: Publisher Site | Google Scholar
  15. D. S. de Jesus, R. J. Cassella, S. L. C. Ferreira, A. C. S. Costa, M. S. de Carvalho, and R. E. Santelli, “Polyurethane foam as a sorbent for continuous flow analysis: pre-concentration and spectrophotometric determination of zinc in biological materials,” Analytica Chimica Acta, vol. 336, no. 1–3, pp. 263–269, 1998. View at: Publisher Site | Google Scholar
  16. D. L. Giokas, E. K. Paleologos, M. I. Prodromidis, and M. I. Karayannis, “Development of 1-(2-pyridylazo)-2-naphthol-modified polymeric membranes for the effective batch pre-concentration and determination of zinc traces with flame atomic absorption spectrometry,” Talanta, vol. 56, no. 3, pp. 491–498, 2002. View at: Publisher Site | Google Scholar
  17. N. Teshima, S. Gotoh, K. Ida, and T. Sakai, “One-shot flow injection spectrophotometric simultaneous determination of copper, iron and zinc in patients' sera with newly developed multi-compartment flow cell,” Analytica Chimica Acta, vol. 557, no. 1-2, pp. 387–392, 2006. View at: Publisher Site | Google Scholar
  18. L. S. G. Teixeira, F. R. P. Rocha, M. Korn, B. F. Reis, S. L. C. Ferreira, and A. C. S. Costa, “Flow-injection solid-phase spectrophotometry for the determination of zinc in pharmaceutical preparations,” Analytica Chimica Acta, vol. 383, no. 3, pp. 309–315, 1999. View at: Publisher Site | Google Scholar
  19. A. C. B. Dias, J. M. T. Carneiro, and E. A. G. Zagatto, “Spectrophotometric flow-injection determination of zinc in plant digests based on a spot test,” Talanta, vol. 63, no. 2, pp. 245–250, 2004. View at: Publisher Site | Google Scholar

Copyright © 2008 Liang Wei et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.


More related articles

 PDF Download Citation Citation
 Download other formatsMore
 Order printed copiesOrder
Views1109
Downloads472
Citations

Related articles

We are committed to sharing findings related to COVID-19 as quickly as possible. We will be providing unlimited waivers of publication charges for accepted research articles as well as case reports and case series related to COVID-19. Review articles are excluded from this waiver policy. Sign up here as a reviewer to help fast-track new submissions.