Trace Lead Measurement and Online Removal of Matrix Interference in Geosamples by Ion-Exchange Coupled with Flow Injection and Hydride Generation Atomic Fluorescence Spectrometry

Table 2

The operating program of the FI ion-exchange system.

StepTimes (s)Pump rate (ml m i n āˆ’ 1 )Valve positionDescription
PaPbPaPb

01801803.500Pa is active, sample load and lead are preconcentration in columns a, and b
111000Pa stops; sample loops dip into water
210103.500Water pushes the remain sample of the loop to pass through the columns
355021Pb is active; remain water of the loop is pushed out by air
411001Pa stops; sample loops dip into the eluent
52020021Pa drives eluent to pass through b and a in series
62020071Pb and Pc are active; sample is reacted with NaBH4 and pushed to AFS for determination.
722400Tube is washed by water