Research Article

An Optical Approach for the Determination of Droplet Volumes in Nanodispensing

Table 7

Gravimetric measurement of different quantities of drops under constant conditions.

Number of drops5000750010 00012 50015 000
Number of measurements1515151515
Mean value [mg or 𝜇 L]0.72 mg1.12 mg1.48 mg1.85 mg2.24 mg
Std. deviation0.03720.03730.03120.03980.0319
% Std. deviation5.2%3.3%2.1%2.2%1.4%
Visual volume200 pl200 pl200 pl200 pl200 pl
Visual/gravimetric ratio1.401.331.351.351.34