Research Article

Determination of Trace Silver in Water Samples by Online Column Preconcentration Flame Atomic Absorption Spectrometry Using Termite Digestion Product

Table 1

Performance data for silver determination by sorption on termite digestion product minicolumn using an FI system.

Parameter

Working range10–50 μg L−1
Correlation coefficient (R)0.99928
Precision (15 μg L−1, 𝑛 = 6 )%3.7
Precision (35 μg L−1, 𝑛 = 6 )%1.1
Limit of detection3.4 μg L−1
Enrichment factor21