Research Article

Bioanalytical Method Development and Validation of Memantine in Human Plasma by High Performance Liquid Chromatography with Tandem Mass Spectrometry: Application to Bioequivalence Study

Table 2

Precision and accuracy (analysis with spiked plasma samples at three different concentrations).

Spiked plasma concentration (pg/ml)Within-run ( 𝑛 = 6 ) Between-run ( 𝑛 = 3 0 )
Concentration measured ( 𝑛 = 6 ) (pg/ml) (mean ± Sd.)Precision (%CV)Accuracy %Concentration measured ( 𝑛 = 3 0 ) (pg/ml) (mean ± Sd.)Precision (%CV)Accuracy %

150.00 1 4 3 . 4 0 ± 3 . 2 0 2.2095.60 1 4 3 . 5 0 ± 3 . 9 0 2.7095.70
15000.00 1 4 7 4 6 . 4 0 ± 3 3 8 . 4 0 2.3098.30 1 4 7 1 9 . 1 0 ± 2 4 8 . 3 0 1.7098.10
35000.00 3 4 9 3 5 . 2 0 ± 7 3 0 . 4 0 2.1099.80 3 4 6 9 9 . 2 0 ± 4 9 8 . 4 0 1.4099.10