Research Article

A Stability-Indicating HPLC-DAD Method for Determination of Stiripentol: Development, Validation, Kinetics, Structure Elucidation and Application to Commercial Dosage Form

Table 2

Intraday ( = 3) and interday ( = 3) precision and accuracy of STP.

Nominal concentration (g/mL)IntradayInterday
Mean ± SD (g/mL)Accuracy (%)RSD (%)Bias (%)Mean ± SD (g/mL)Accuracy (%)RSD (%)Bias (%)

22.01 ± 0.04100.682.030.51.95 ± 0.0997.781.41−2.5
66.10 ± 0.04101.640.641.676.02 ± 0.0799.901.180.33
109.87 ± 0.1898.741.76−1.39.99 ± 0.4099.752.65−0.1
1514.85 ± 0.3598.992.32−1.014.95 ± 0.4299.852.32−0.3
2020.07 ± 0.39100.361.940.3519.95 ± 0.5499.553.03−0.2