Journal of Analytical Methods in Chemistry
Volume 2014, Article ID 984314, 10 pages
http://dx.doi.org/10.1155/2014/984314
Research Article

Simultaneous Determination of Uric Acid and Xanthine Using a Poly(Methylene Blue) and Electrochemically Reduced Graphene Oxide Composite Film Modified Electrode

Department of Chemistry and Materials Science, Huaibei Normal University, Huaibei 235000, China

Received 29 July 2014; Revised 9 October 2014; Accepted 19 October 2014; Published 11 November 2014

Academic Editor: Chih-Ching Huang

Copyright © 2014 Gen Liu et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

Poly(methylene blue) and electrochemically reduced graphene oxide composite film modified electrode (PMB-ERGO/GCE) was successfully fabricated by electropolymerization and was used for simultaneous determination of uric acid (UA) and xanthine (Xa). Based on the excellent electrocatalytic activity of PMB-ERGO/GCE, the electrochemical behaviors of UA and Xa were studied by cyclic voltammetry (CV) and square wave voltammetry (SWV). Two anodic sensitive peaks at 0.630 V (versus Ag/AgCl) for UA and 1.006 V (versus Ag/AgCl) for Xa were given by CV in pH 3.0 phosphate buffer. The calibration curves for UA and Xa were obtained in the range of 8.00 × 10−8~4.00 × 10−4 M and 1.00 × 10−7~4.00 × 10−4 M, respectively, by SWV. The detection limits for UA and Xa were  M and  M, respectively. Finally, the proposed method was applied to simultaneously determine UA and Xa in human urine with good selectivity and high sensitivity.

1. Introduction

Graphene oxide (GO) and graphene are promising materials with many interesting properties for applications [1]; for example, they have nice compatibility with biological materials such as amino acids [2] and proteins [3]. Although its badly electrical conductivity limits the usage of graphene oxide, many efforts have improved the methods to obtain the reduced graphene oxide, such as chemical reduced [4, 5], UV-induced photocatalytic reduced [6], and electrochemical reduced [7]. Electrochemically reduced graphene oxide could be prepared based on the research of Zhou et al. [8]. Moreover, Casero et al. immersed a graphite working electrode in GO dispersion and controlled an appropriate cathodic potential in 12 hours to obtain reduced graphene oxide [9].

The electron pair or electron cloud overlapping between dye molecules and graphite electrode strengthened their adsorption affinity and accelerated electron transfer rate [10]. Because of the π-π noncovalent interaction, some electroactive water-soluble aromatic dyes could absorb on the surface of carbon material, such as the combination of Toluidine blue O and multiwall carbon nanotube [11]. Methylene blue (MB), one of the phenoxazine biological dyes, is an electron carrier activation. Sun et al. had applied poly(methylene blue) functionalized graphene modified carbon ionic liquid electrode to determine dopamine. The modified electrode exhibited better electrochemical performances with higher conductivity and lower electron transfer resistance [12]. It indicates that poly(methylene blue) functionalized graphene possesses distinct properties in electrochemical field.

Uric acid (UA, C5H4N4O3) is one of the final metabolites of proteins and nucleic acids. It is mainly excreted by kidney and to a less extent by liver. To most mammals and birds, UA is biologically decomposed to allantoin by the action of uricase and then decomposed to NH3, CO2, and H2O. However, the further step is not applicable in humans and apes because of uricase insufficiency. Consequently, UA is the final metabolic product of purine. The amount of UA in human body has great clinical values since it has a bearing in the diagnosis of gout, Lesch-Nyhan syndrome, urolithiasis, kidney damage, leukemia, lymphoma [1315], and so forth.

Xanthine (Xa, C5H4N4O2) mainly exists in the animals’ blood, liver, and urine and is an important metabolic intermediate of the purine nucleotide and deoxynucleotide. In addition, the decomposition of adenosine triphosphate (ATP) can also produce Xa [16]. In clinical diagnosis, Xa usually acts as a sensitive indicator for some clinical disorders such as perinatal asphyxia and adult respiratory distress syndrome [1719]. As the metabolic precursor of UA, Xa determines the level of UA in body fluid. Thus, developing a stable, sensitive, and selective sensor for UA and Xa has considerable importance for clinical analysis. Previously, various methods for the determination of UA and Xa have been reported, such as fluorimetry [20, 21], spectrophotometry [22, 23], enzyme assay [15, 24], chromatography [25, 26], and electrochemical methods [2729]. However, some inherent problems limit their applications, which includes expensive instruments, complex operations, and personnel training.

In this work, we fabricated poly(methylene blue) and electrochemically reduced graphene oxide composite film modified electrode, which comprised not only excellent electrocatalytic activity for the oxidation of UA and Xa, but also good sensitivity, wide linearity, high selectivity, and remarkable reproducibility. Thus, this modified electrode was satisfactorily used for the simultaneous determination of UA and Xa in human urine by square wave voltammetry (SWV).

2. Experimental

2.1. Instruments and Chemicals

Electrochemical measurements were performed with a BAS 100 B/W electrochemical workstation (BAS, USA). A conventional three-electrode system was used for all electrochemical experiments, which consisted of an Ag/AgCl (saturated KCl) as the reference electrode, a platinum wire as the auxiliary electrode, and a bare GCE (3 mm in radius) or modified GCE as working electrode. The morphology of the samples was observed using scanning electron microscopes (SEM) from JSM-6610LV (JEOL, Japan). The UV-visible spectra were obtained by UV-3600 UV-VIS-NIR spectrophotometer (Shimadzu, Japan). All pH measurements were performed using a PHS-3C digital pH meter (Shanghai Leici Device Works, China).

Graphene oxide dispersion (2 mg·mL−1; dispersing agent: water; radius: 1~5 μm; thickness: 0.8~1.2 nm; single layer ratio: 99%; purity: 99%) was obtained from Nanjing XFNANO Materials Technology Company, China. MB, UA, and Xa were purchased from Sigma-Aldrich.All other chemicals were used of analytical grade. Double distilled water was used throughout the experiments. The pH of phosphate buffered solution (PBS) was adjusted with 0.1 M H3PO4 and 0.1 M NaOH. All experiments were carried out at room temperature.

2.2. Preparation of PMB-ERGO/GCE

The bare GCE was polished with 0.05 μm α-alumina powder, until a mirror-like finish, rinsed with 1 : 1 HNO3 solution, ethanol, and doubly distilled water in an ultrasonic aqueous bath successively, and again rinsed with water before it was used. 1.00 mg·mL−1 GO in 0.1 M PBS (pH 6.0) was ultrasonicated for 5 min; then an appropriate amount of MB was added to form an aqueous solution (2.00 × 10−3 M), followed by ultrasonication for another 5 min. Finally, the polymeric film was electropolymerized by sweeping from 0.7 to −1.2 V at a scan rate of 0.10 V·s−1 for 11 cycles using cyclic voltammetry. After polymerization, the modified electrode was removed, rinsed thoroughly with double distilled water, and air-dried. Finally the poly(methylene blue) and electrochemically reduced graphene oxide composite film modified electrode (PMB-ERGO/GCE) was successfully prepared.

3. Results and Discussion

3.1. Characterizations of PMB-ERGO/GCE

Compared with lower and higher electropolymerized cathodic potential, potential of −1.2 V was enough to guarantee good catalytic ability of PMB-ERGO/GCE toward UA and Xa. Besides, lower cathodic potential usually produced hydrogen bubbles on the surface of electrodes without further improving the responses to UA and Xa. In fact, these bubbles not only influenced the electric current density, but also damaged the modified film.

Figure 1(a) gives the cyclic voltammograms of electrochemical polymerization for poly(methylene blue) film modified electrode (PMB/GCE), and Figure 1(b) is PMB-ERGO/GCE. It can be seen in Figure 1(a) that a pair of sharp redox peaks of MB appear at 0.162 V (anodic peak) and −0.181 V (cathodic peak) in the first cycle. This redox reaction possesses poor reversibility in that a large and obvious difference exists in the redox peak currents. In Figure 1(b), however, the anodic peak (at 0.151 V) and the cathodic peak (at 0.022 V) of MB are both wide; moreover, GO has an irreversible cathodic peak at −1.079 V. In the following cycles, the cathodic peak current of GO decreases, the redox peak currents of MB increase, and finally these curves are not changed because of the increasing thickness of modified film and the deceleration of the rate of deposition.

fig1
Figure 1: CVs of PMB/GCE (a) and PMB-ERGO/GCE (b) in the polymerization process at a scan rate of 0.10 V·s−1 for 11 cycles (from a to k); : anodic peak of MB; : cathodic peak of MB; : cathodic peak of GO.

It is well-known that some oxygen-containing functional groups, such as –OH, –COOH, –C=O, –C–O–C–, which exist in graphene oxide, can lead to the large amount of sp3-hybridized carbon atoms. Electrochemical reduction is a technique to remove these oxygen functional groups and can restore sp2-hybridized carbon atoms, which can change graphene oxide to graphene. After polymerization, the electrochemically reduced graphene oxide (ERGO) is prepared on the surface of electrodes.

Figure 2 exhibits UV-visible spectra curves of 1.50 × 10−5 M MB and 0.02 mg·mL−1 GO and the mixture solution of 1.50 × 10−5 M MB and 0.02 mg·mL−1 GO. MB gives two absorption peaks at 662 nm (1) and 616 nm (2), and peak 1 is much spikier and stronger than peak 2. The absorption peaks might originate from the corresponding monomer and dimer of MB, respectively [35]. GO has no obvious absorption. After GO is added to MB solution, the absorption peaks of MB decline drastically, but still at the same wavelength, implying that some interactions occur between MB and GO.

984314.fig.002
Figure 2: Comparison of UV-visible spectra of MB (a), GO (b), and the mixture solution of MB and GO (c) in 1.0 cm light path length cell.

SEM images of GCE, PMB/GCE, and PMB-ERGO/GCE are displayed in Figure 3. PMB film is smooth, dense, and uniform and PMB-ERGO film shows a rough surface feature with bulges, which indicates that ERGO is embedded into the polymer structure of MB and increases the specific surface area. Compared with bare GCE, the electropolymerized films of PMB and PMB-ERGO were colorful and could be observed by naked eye, suggesting the films had been successfully adhered to the electrode surface.

984314.fig.003
Figure 3: SEM images of GCE, PMB/GCE, and PMB-ERGO/GCE.

Figure 4 depicts the electrochemical impedance spectra of GCE, PMB/GCE, ERGO/GCE, and PMB-ERGO/GCE, and their corresponding impedance values were 1212 Ω, 233.1 Ω, 96.4 Ω, and 162.5 Ω, respectively. These data imply that the modified films of PMB, ERGO, and PMB-ERGO had accelerated the rate and lowered the activation energy of electron transfer, particularly ERGO.

984314.fig.004
Figure 4: EIS of GCE, PMB/GCE, ERGO/GCE, and PMB-ERGO/GCE in 5.0 × 10−3 M K3[Fe(CN)6], 1.0 M KCl solution with initial for 150 mV, high frequency for 2000 Hz, low frequency for 0.05 Hz, and A.C. amplitude for 150 mV.
3.2. Electrocatalytic Activity of UA and Xa at Different Electrodes

Figure 5 shows the CV responses of the bare GCE, PMB/GCE, ERGO/GCE, and PMB-ERGO/GCE toward 2.00 × 10−4 M UA and 2.00 × 10−4 M Xa. Table 1 lists the anodic peak currents and potentials of different electrodes. It is noticed that the anodic peak potentials on the four electrodes are close. However, the anodic peak currents of UA and Xa on PMB-ERGO/GCE are much higher than others, indicating PMB-ERGO/GCE has the greatest electrochemical activity, which ought to be attributed to the synergistic effects of PMB and ERGO. On PMB-ERGO/GCE, Epa(UA) = 0.630 V, and Epa(Xa) = 1.006 V, their peaks were separated by 0.376 V. This result indicates that the simultaneous determination for UA and Xa could be achieved without separation. Combining the nice conductivity, PMB-ERGO/GCE is appropriate to be employed to investigate the electrochemical behavior of UA and Xa.

tab1
Table 1: Detail data of cyclic voltammograms at different electrodes.
984314.fig.005
Figure 5: CVs of bare GCE, PMB/GCE, ERGO/GCE, and PMB-ERGO/GCE in the mixture solution of UA and Xa and PBS (pH 3.0) at a scan rate of 0.14 V·s−1 with quiet time for 2 min.
3.3. Effective Area of Electrodes

GCE, PMB/GCE, ERGO/GCE, and PMB-ERGO/GCE were immerged into the solution containing 5.0 × 10−3 M K3[Fe(CN)6] and 1.0 M KCl, followed by sweeping with cyclic voltammetry. The effective area of electrode can be calculated from the Randles-Sevcik plot: , wherein ip is current in amps (A), n is number of electrons transferred of K3[Fe(CN)6] in the redox event (usually is 1), A is electrode effective area, D is diffusion coefficient (7.6 × 10−6 cm2·s−1), C is concentration (5 mM), and ν is scan rate (0.05 V·s−1). The effective areas for GCE, PMB/GCE, ERGO/GCE, and PMB-ERGO/GCE were 0.0941 cm2, 0.1906 cm2, 0.2088 cm2, and 0.2291 cm2, respectively. PMB-ERGO/GCE contains the largest effective area, which presumably accounts for its excellent electrocatalytic activity.

3.4. Effect of pH

In most cases, the electrolyte pH is an important condition to the electrochemical reaction. The cyclic voltammograms of 5.00 × 10−5 M UA and 1.00 × 10−4 M Xa were recorded from pH 2.0 to 8.0. As shown in Figure 6, the anodic peak potentials shift in negative direction with a rising value of pH, suggesting that protons have participated in electrode reactions. The anodic peak current of UA increases up to pH 3.0 and then decreases. However, the anodic peak current of Xa decreases in response to increasing pH. Table 2 lists the relationship between peak potential and pH of UA and Xa. Based on the analysis above, pH 3.0 and pH 2.0 were used in individual determination for UA and Xa, respectively, and pH 3.0 was used in their simultaneous determination.

tab2
Table 2: The relationship between peak potential and pH of UA and Xa.
fig6
Figure 6: CVs of UA (a) and Xa (b) with different pH. Insets show the calibration plots of E-pH. From a to m: pH = 2.0, 2.5, 3.0, 3.5, 4.0, 4.5, 5.0, 5.5, 6.0, 6.5, 7.0, 7.5, 8.0.
3.5. Effect of Scan Rates

To investigate the effect of scan rate on the electrochemical behavior of UA and Xa (2.50 × 10−5 M) on PMB-ERGO/GCE, cyclic voltammograms were performed by various potential scan rates. Figure 7 shows the anodic peak potential shifts to positive region with increasing scan rates; in addition, the anodic peak current increases at the same time. The relationships between scan rate, peak current, and peak potential are expressed with equations in Table 3.

tab3
Table 3: The relationship between scan rate, peak current, and peak potential.
fig7
Figure 7: CVs for UA (a) and Xa (b) at different scan rate in PBS (pH 3.0). Insets show the calibration plots of lgI-lgv and E-lnv. From a to m: 0.04, 0.06, 0.08, 0.10, 0.12, 0.14, 0.16, 0.18, 0.20, 0.22, 0.24, 0.26, 0.28.

The slope values approximate to 0.5 in equations of lgI-lgv, which clearly reveals that the electron transfer reactions of UA and Xa are both controlled by diffusion. Formula RT/(αnF) is usually used to estimate the slope in E-lnv relationship [36]; herein, n is the reaction electron number, F is Faraday constant, R is gas constant, and α represents electron transfer coefficient (usually is 0.5). The real reaction electron numbers of UA and Xa were calculated to be 1.964 and 1.913, respectively. Their theoretical reaction electron numbers should be 2. According to literatures [27, 37, 38], the mechanisms of UA and Xa oxidation can be proposed as follows in Figures 8 and 9, respectively.

984314.fig.008
Figure 8: The electrochemical reaction mechanism for UA.
984314.fig.009
Figure 9: The electrochemical reaction mechanism for Xa.
3.6. Individual and Simultaneous Determination of UA and Xa

Figures 10 and 11 describe CV and SWV responses of UA, Xa, and their mixture solution at various concentrations, respectively. And Table 4 lists their linear ranges, detection limits, and so forth.

tab4
Table 4: Analytical parameters for individual and simultaneous determination of UA and Xa at PMB-ERGO/GCE.
fig10
Figure 10: CVs of different concentrations of UA (a), Xa (b), and the mixture solution of UA and Xa (c) (from a to k: 0.08, 0.1, 0.5, 1, 2.5, 5, 7.5, 10, 25, 50, 75, 100, 200, 300, 400 μM) at a scan rate of 0.14 V·s−1 with quiet time for 2 min. Insets show the calibration plots of lgI-lgC.
fig11
Figure 11: SWVs of different concentrations of UA (a) (from a to p: 0.08, 0.1, 0.5, 1, 2.5, 5, 7.5, 10, 25, 50, 75, 100, 200, 300, 400, 500 μM), Xa (b) (from a to p: 0.05, 0.08, 0.1, 0.5, 1, 2.5, 5, 7.5, 10, 25, 50, 75, 100, 200, 300, 400 μM), and the mixture solution of UA and Xa (c) (from a to o: 0.08, 0.1, 0.5, 1, 2.5, 5, 7.5, 10, 25, 50, 75, 100, 200, 300, 400 μM) with step for 5 mV, amplitude for 30 mV, frequency for 5 Hz, and quiet time for 2 min. Insets show the calibration plots of lgI-lgC.

The results in Table 4 demonstrated that the proposed method had wide linear range and high sensitivity. The comparison between this method and other electrochemical methods for simultaneous determination of UA and Xa was listed in Table 5.

tab5
Table 5: Comparison of the proposed method with other electrochemical methods for the simultaneous determination of UA and Xa.
3.7. Reproducibility and Stability

The mixed solution of 5.00 × 10−5 M UA and 1.00 × 10−4M Xa were SWV measured for 20 parallel experiments. The relative standard deviations (RSD) of the peak currents for UA and Xa were 3.1% and 2.7%, respectively, implying remarkable reproducibility. After the modified electrode was stored in humid environment at room temperature for 15 days, it retained 94.5% of its original response and held the similar shape of the original curves, suggesting an acceptable stability of PMB-ERGO/GCE.

3.8. Interference

The influences of various foreign species were investigated in a mixture solution containing 2.00 × 10−5M UA and 5.00 × 10−5M Xa. The tolerance limit was set as the maximum concentration of the foreign substances that caused an approximately ±5% relative error in the determination. The results showed that K+, Na+, Ca2+, Fe2+, Mg2+, Ba2+, Zn2+, Al3+, Cl, , , , starch, L-arginine, L-threonine, L-serine, L-histidine, dopamine (≥1.0mg) and L-cysteine (0.6mg), Cu2+ (0.6mg), Ag+ (0.05mg), I (0.03mg), and ascorbic acid (0.5mg) had no interference with the determination of UA and Xa. Therefore, it is possible to simultaneously determine UA and Xa in a sample on PMB-ERGO/GCE.

3.9. Sample Analysis

Human urine sample was selected as the real sample to examine the reliability of the proposed SWV method. 1 mL urine sample was diluted 10 times with 0.1M PBS (pH 3.0) before measurement. The obtained results are summarized in Table 6.

tab6
Table 6: Simultaneous determination of UA and Xa in human urine samples ().

4. Conclusions

PMB-ERGO/GCE was prepared by electropolymerization. The modified electrode exhibited good conductivity and excellent electrocatalytic activity toward UA and Xa. The introduction of ERGO enhanced the effective surface areas on modified electrode compared with PMB film. The real reaction electron number of UA and Xa were calculated to be 1.964 and 1.913, respectively, in pH 3.0 phosphate buffer solution. The oxidation of UA and Xa were both controlled by diffusion. PMB-ERGO/GCE displayed desirable properties including excellent stability, reproducibility, selectivity, and sensitivity. The results demonstrated that the proposed method is a rapid, sensitive, and reproducible method for determination of UA and Xa in human urine sample. Therefore, PMB-ERGO/GCE would act as a promising sensor for a wide range of electrochemical sensing and biosensing applications.

Conflict of Interests

The authors declare that there is no conflict of interests regarding the publication of this paper.

Acknowledgment

This research is funded by Anhui Province Universities Provincial Natural Science Research Foundation for Key Project (no. KJ2011A255), China.

References

  1. C. Y. Liu, Z. Y. Liu, R. Peng, and Z. C. Zhong, “Quasireversible process of dopamine on copper-nickel hydroxide composite/nitrogen doped graphene/nafion modified GCE and its electrochemical application,” Journal of Analytical Methods in Chemistry, vol. 2014, Article ID 724538, 7 pages, 2014. View at Publisher · View at Google Scholar
  2. L. Hua, X. Wu, and R. Wang, “Glucose sensor based on an electrochemical reduced graphene oxide-poly(l-lysine) composite film modified GC electrode,” Analyst, vol. 137, no. 24, pp. 5716–5719, 2012. View at Publisher · View at Google Scholar · View at Scopus
  3. X. Zuo, S. He, D. Li et al., “Graphene oxide-facilitated electron transfer of metalloproteins at electrode surfaces,” Langmuir, vol. 26, no. 3, pp. 1936–1939, 2010. View at Publisher · View at Google Scholar · View at Scopus
  4. S. Pei, J. Zhao, J. Du, W. Ren, and H.-M. Cheng, “Direct reduction of graphene oxide films into highly conductive and flexible graphene films by hydrohalic acids,” Carbon, vol. 48, no. 15, pp. 4466–4474, 2010. View at Publisher · View at Google Scholar · View at Scopus
  5. W. Guoxiu, Y. Juan, P. Jinsoo et al., “Facile synthesis and characterization of graphene nanosheets,” Journal of Physical Chemistry C, vol. 112, no. 22, pp. 8192–8195, 2008. View at Publisher · View at Google Scholar · View at Scopus
  6. G. Williams, B. Seger, and P. V. Kamt, “TiO2-graphene nanocomposites. UV-assisted photocatalytic reduction of graphene oxide,” ACS Nano, vol. 2, no. 7, pp. 1487–1491, 2008. View at Publisher · View at Google Scholar · View at Scopus
  7. X. Zhang, D. c. Zhang, Y. Chen, X. Z. Sun, and Y. W. Ma, “Electrochemical reduction of graphene oxide films: preparation, characterization and their electrochemical properties,” Chinese Science Bulletin, vol. 57, no. 23, pp. 3045–3050, 2012. View at Publisher · View at Google Scholar · View at Scopus
  8. M. Zhou, Y. L. Wang, Y. M. Zhai et al., “Controlled synthesis of large-area and patterned electrochemically reduced graphene oxide films,” Chemistry, vol. 15, no. 25, pp. 6116–6120, 2009. View at Publisher · View at Google Scholar · View at Scopus
  9. E. Casero, C. Alonso, L. Vázquez et al., “Comparative response of biosensing platforms based on synthesized graphene oxide and electrochemically reduced graphene,” Electroanalysis, vol. 25, no. 1, pp. 154–165, 2013. View at Publisher · View at Google Scholar · View at Scopus
  10. L. Gorton, G. Johansson, and A. Torstensson, “A kinetic study of the reaction between dihydronicotinamide adenine dinucleotide (NADH) and an electrode modified by adsorption of 1,2-benzophenoxazine-7-one,” Journal of Electroanalytical Chemistry, vol. 196, no. 1, pp. 81–92, 1985. View at Publisher · View at Google Scholar · View at Scopus
  11. J. X. Zeng, W. Z. Wei, L. Wu, X. Y. Liu, K. Liu, and Y. Li, “Fabrication of poly(toluidine blue O)/carbon nanotube composite nanowires and its stable low-potential detection of NADH,” Journal of Electroanalytical Chemistry, vol. 595, no. 2, pp. 152–160, 2006. View at Publisher · View at Google Scholar
  12. W. Sun, Y. Wang, Y. Zhang, X. Ju, G. Li, and Z. Sun, “Poly(methylene blue) functionalized graphene modified carbon ionic liquid electrode for the electrochemical detection of dopamine,” Analytica Chimica Acta, vol. 751, pp. 59–65, 2012. View at Publisher · View at Google Scholar · View at Scopus
  13. V. V. S. Eswara Dutt and H. A. Mottola, “Determination of uric acid at the microgram level by a kinetic procedure based on a “pseudo-induction” period,” Analytical Chemistry, vol. 46, no. 12, pp. 1777–1781, 1974. View at Publisher · View at Google Scholar · View at Scopus
  14. X. Chen, G. Wu, Z. Cai, M. Oyama, and X. Chen, “Advances in enzyme-free electrochemical sensors for hydrogen peroxide, glucose, and uric acid,” Microchimica Acta, vol. 181, no. 7-8, pp. 689–705, 2014. View at Publisher · View at Google Scholar · View at Scopus
  15. J. Yu, S. Wang, L. Ge, and S. Ge, “A novel chemiluminescence paper microfluidic biosensor based on enzymatic reaction for uric acid determination,” Biosensors and Bioelectronics, vol. 26, no. 7, pp. 3284–3289, 2011. View at Publisher · View at Google Scholar · View at Scopus
  16. X. Tang, Y. Liu, H. Hou, and T. You, “A nonenzymatic sensor for xanthine based on electrospun carbon nanofibers modified electrode,” Talanta, vol. 83, no. 5, pp. 1410–1414, 2011. View at Publisher · View at Google Scholar · View at Scopus
  17. S. Z. Bas, E. Maltas, B. Sennik, and F. Yilmaz, “Design of an electrochemical biosensing system for xanthine detection and a study on binding interaction of ketoconazole with xanthine oxidase,” Colloids and Surfaces A: Physicochemical and Engineering Aspects, vol. 444, pp. 40–47, 2014. View at Publisher · View at Google Scholar · View at Scopus
  18. X. Liu, W. M. Lin, X. H. Yan, X. H. Chen, J. R. Hoidal, and P. Xu, “Improved method for measurement of human plasma xanthine oxidoreductase activity,” Journal of Chromatography B: Analytical Technologies in the Biomedical and Life Sciences, vol. 785, no. 1, pp. 101–114, 2003. View at Publisher · View at Google Scholar · View at Scopus
  19. M. Amiri-Aref, J. B. Raoof, and R. Ojani, “A highly sensitive electrochemical sensor for simultaneous voltammetric determination of noradrenaline, acetaminophen, xanthine and caffeine based on a flavonoid nanostructured modified glassy carbon electrode,” Sensors and Actuators B: Chemical, vol. 192, pp. 634–641, 2014. View at Publisher · View at Google Scholar · View at Scopus
  20. J. Galban, Y. Andreu, M. J. Almenara, S. de Marcos, and J. R. Castillo, “Direct determination of uric acid in serum by a fluorometric-enzymatic method based on uricase,” Talanta, vol. 54, no. 5, pp. 847–854, 2001. View at Publisher · View at Google Scholar · View at Scopus
  21. D. Martinez-Pérez, M. L. Ferrer, and C. R. Mateo, “A reagent less fluorescent sol-gel biosensor for uric acid detection in biological fluids,” Analytical Biochemistry, vol. 322, no. 2, pp. 238–242, 2003. View at Publisher · View at Google Scholar · View at Scopus
  22. D. L. Rocha and F. R. P. Rocha, “A flow-based procedure with solenoid micro-pumps for the spectrophotometric determination of uric acid in urine,” Microchemical Journal, vol. 94, no. 1, pp. 53–59, 2010. View at Publisher · View at Google Scholar · View at Scopus
  23. X.-X. Wang, Q. Wu, Z. Shan, and Q.-M. Huang, “BSA-stabilized Au clusters as peroxidase mimetics for use in xanthine detection,” Biosensors and Bioelectronics, vol. 26, no. 8, pp. 3614–3619, 2011. View at Publisher · View at Google Scholar · View at Scopus
  24. E. E. Kelley, A. Trostchansky, H. Rubbo, B. A. Freeman, R. Radi, and M. M. Tarpey, “Binding of xanthine oxidase to glycosaminoglycans limits inhibition by oxypurinol,” The Journal of Biological Chemistry, vol. 279, no. 36, pp. 37231–37234, 2004. View at Publisher · View at Google Scholar · View at Scopus
  25. Y. Zuo, Y. Yang, Z. Zhu, W. He, and Z. Aydin, “Determination of uric acid and creatinine in human urine using hydrophilic interaction chromatography,” Talanta, vol. 83, no. 5, pp. 1707–1710, 2011. View at Publisher · View at Google Scholar · View at Scopus
  26. T. Yamamoto, Y. Moriwaki, S. Takahashi et al., “Determination of human plasma xanthine oxidase activity by high-performance liquid chromatography,” Journal of Chromatography B: Biomedical Applications, vol. 681, no. 2, pp. 395–400, 1996. View at Publisher · View at Google Scholar · View at Scopus
  27. X. Q. Tian, C. M. Cheng, H. Y. Yuan et al., “Simultaneous determination of L-ascorbic acid, dopamine and uric acid with gold nanoparticles-β-cyclodextrin-graphene-modified electrode by square wave voltammetry,” Talanta, vol. 93, pp. 79–85, 2012. View at Publisher · View at Google Scholar · View at Scopus
  28. M. Mohamadi, A. Mostafavi, and M. Torkzadeh-Mahani, “Voltammetric behavior of uric acid on carbon paste electrode modified with salmon sperm dsDNA and its application as label-free electrochemical sensor,” Biosensors and Bioelectronics, vol. 54, pp. 211–216, 2014. View at Publisher · View at Google Scholar · View at Scopus
  29. R. Devi, S. Yadav, and C. S. Pundir, “Au-colloids-polypyrrole nanocomposite film based xanthine biosensor,” Colloids and Surfaces A: Physicochemical and Engineering Aspects, vol. 394, pp. 38–45, 2012. View at Publisher · View at Google Scholar · View at Scopus
  30. P. Kalimuthu and S. A. John, “Simultaneous determination of ascorbic acid, dopamine, uric acid and xanthine using a nanostructured polymer film modified electrode,” Talanta, vol. 80, no. 5, pp. 1686–1691, 2010. View at Publisher · View at Google Scholar · View at Scopus
  31. Y. Wang and L.-L. Tong, “Electrochemical sensor for simultaneous determination of uric acid, xanthine and hypoxanthine based on poly (bromocresol purple) modified glassy carbon electrode,” Sensors and Actuators B: Chemical, vol. 150, no. 1, pp. 43–49, 2010. View at Publisher · View at Google Scholar · View at Scopus
  32. L. Yang, D. Liu, J. Huang, and T. You, “Simultaneous determination of dopamine, ascorbic acid and uric acid at electrochemically reduced graphene oxide modified electrode,” Sensors and Actuators B: Chemical, vol. 193, pp. 166–172, 2014. View at Publisher · View at Google Scholar · View at Scopus
  33. J.-M. Zen, Y.-Y. Lai, H.-H. Yang, and A. S. Kumar, “Multianalyte sensor for the simultaneous determination of hypoxanthine, xanthine and uric acid based on a preanodized nontronite-coated screen-printed electrode,” Sensors and Actuators B: Chemical, vol. 84, no. 2-3, pp. 237–244, 2002. View at Publisher · View at Google Scholar · View at Scopus
  34. F. Y. Zhang, Z. H. Wang, Y. Z. Zhang et al., “Simultaneous electrochemical determination of uric acid, xanthine and hypoxanthine based on poly(L-arginine)/graphene composite film modified electrode,” Talanta, vol. 93, pp. 320–325, 2012. View at Publisher · View at Google Scholar
  35. L.-X. Sun, A. M. Reddy, N. Matsuda, A. Takatsu, K. Kato, and T. Okada, “Simultaneous determination of methylene blue and new methylene blue by slab optical waveguide spectroscopy and artificial neural networks,” Analytica Chimica Acta, vol. 487, no. 1, pp. 109–116, 2003. View at Publisher · View at Google Scholar · View at Scopus
  36. E. Laviron, “Adsorption, autoinhibition and autocatalysis in polarography and in linear potential sweep voltammetry,” Journal of Electroanalytical Chemistry, vol. 52, no. 3, pp. 355–393, 1974. View at Publisher · View at Google Scholar · View at Scopus
  37. G. Dryhurst, “Electrochemical oxidation of uric acid and xanthine at the pyrolytic graphite electrode: mechanistic interpretation of electrochemistry,,” Journal of the Electrochemical Society, vol. 119, no. 12, pp. 1659–1664, 1972. View at Publisher · View at Google Scholar · View at Scopus
  38. Z. H. Wang, S. F. Xiao, and Y. Chen, “β-Cyclodextrin incorporated carbon nanotubes-modified electrodes for simultaneous determination of adenine and guanine,” Journal of Electroanalytical Chemistry, vol. 589, no. 2, pp. 237–242, 2006. View at Publisher · View at Google Scholar · View at Scopus