Research Article

A Stability-Indicating HPLC-DAD Method for Determination of Ferulic Acid into Microparticles: Development, Validation, Forced Degradation, and Encapsulation Efficiency

Table 5

Robustness data for ferulic acid analysis ().

ParameterDrug content (%) ± SDRSDRetention time (min)

Flow rate (mL/min)
 0.9102.71 ± 1.931.885.39
 1.0100.12 ± 0.600.604.86
 1.199.78 ± 0.570.574.49
pH of mobile phase
 3.0100.12 ± 0.600.604.86
 4.097.56 ± 0.981.004.80
 6.098.03 ± 0.370.384.52
Temperature (°C)
 Ambient100.12 ± 0.600.604.86
 30100.09 ± 1.091.094.71
 4099.14 ± 2.222.244.44

SD = standard deviation; RSD = relative standard deviation.