Results of the system suitability test for LER enantiomer analysis.

ParameterUSP recommendationDay(S-LER)(R-LER)

Migration timeRSDa (%) <210.070.07
20.30.3

Plates numberValues >20001357 475328 250
2354 674324 078

AsymmetryValues <211.21.2
21.21.2

ResolutionValues >212.0
22.0

Purity peak Values 110.90.9

Values from five replicates analyses, RSD (%), relative standard deviation expressed as a percentage, and values almost equal to.