Research Article

Development and Validation of a Chromatography Method Using Tandem UV/Charged Aerosol Detector for Simultaneous Determination of Amlodipine Besylate and Olmesartan Medoxomil: Application to Drug-Excipient Compatibility Study

Table 1

Linear and polynomial fit equations for calibration curves for AMLO and OLME determined by HPLC-DAD and HPLC-CAD.

Range (μg mL−1)EquationR2LOD (μg mL−1)LOQ (μg mL−1)

AMLO15–250DAD-linear0.995314
y = 0.5862x + 1.4772
CAD-linear0.0083410
y = 0.0221x − 0.0356
CAD-polynomial0.9994410
y = −2.10−5x2 + 0.0271x + 0.1454

OLME25–500DAD-linear0.9958110
y = 0.6918x + 4.1201
CAD-linear0.9908515
y = 0.0159x − 0.817
CAD-polynomial0.9988515
y = −2.10−5x2 + 0.0257x + 0.2194