Research Article

Development and Validation of a Chromatography Method Using Tandem UV/Charged Aerosol Detector for Simultaneous Determination of Amlodipine Besylate and Olmesartan Medoxomil: Application to Drug-Excipient Compatibility Study

Table 2

Precision and accuracy for AMLO and OLME determined by HPLC-DAD and HPLC-CAD.

Concentration added (μg mL−1)DAD (n = 5)CAD (n = 5)
Recovery (%)RSD intraday (%)RSD interday (%)Recovery (%)RSD intraday (%)RSD interday (%)

AMLO2095.460.540.9695.520.570.88
40101.850.560.81100.720.490.85
80100.310.420.6899.450.510.91

OLME5099.340.480.8999.220.520.93
100100.660.450.87100.890.430.84
15097.920.390.6799.720.420.89

n: number of replicates.