Research Article

Chromatographic Characterization and Method Development for Determination of Levetiracetam in Saliva: Application to Correlation with Plasma Levels

Table 2

Comparison of analytical performance of the proposed method with relevant published methods.

Validation parameterProposed methodGuo et al., 2007
LC-MS/MS
Mecarelli et al., 2007
GC-MS

LOQ (µg/ml)0.2411.1
Linearity ()>0.992NA0.998
Precision (RSD)7.6–4.3%8.2–5.3%13–2%
Recovery98.2–99.6103–108NA
Accuracy (% error)4.2–1.6%Judged by recovery17.9–0.3%

Lowest concentration on calibration curve.