A Standardized Approach to Quantitative Analysis of Nicotine in e-Liquids Based on Peak Purity Criteria Using High-Performance Liquid Chromatography
Table 7
Accuracy results of the method.
| % spiked level of assay sample | Replicate | % recovery | % mean recovery | % RSD |
| 50 (40 µg/mL) | 1 | 99.95 | 99.44 (39.78 µg/mL) | 0.77 | 2 | 99.81 | 3 | 98.56 |
| 100 (80 µg/mL) | 1 | 99.41 | 100.01 (80.01 µg/mL) | 0.95 | 2 | 99.52 | 3 | 101.11 |
| 150 (120 µg/mL) | 1 | 100.27 | 100.61 (120.73 µg/mL) | 0.39 | 2 | 101.04 | 3 | 100.51 |
|
|