Journal of Analytical Methods in Chemistry / 2018 / Article / Tab 5 / Research Article
A Stability Indicating HPLC Assay Method for Analysis of Rivastigmine Hydrogen Tartrate in Dual-Ligand Nanoparticle Formulation Matrices and Cell Transport Medium Table 5 Accuracy data for RHT in NPs matrix, release medium, and cell transport medium.
Prepared RHT concentration (µ g/mL) RHT in NPs matrix RHT in release medium RHT in cell transport medium Measured concentration (µ g/mL) Recovery (%) Measured concentration (µ g/mL) Recovery (%) Measured concentration (µ g/mL) Recovery (%) 0.50 0.49 ± 0.13 98.0 ± 0.2 0.49 ± 0.11 98.4 ± 0.1 0.49 ± 0.16 98.6 ± 0.2 2.00 1.99 ± 0.16 99.5 ± 0.3 2.01 ± 0.10 100.5 ± 0.3 2.00 ± 0.12 99.8 ± 0.2 4.00 3.96 ± 0.19 99.0 ± 0.3 4.04 ± 0.24 101.0 ± 0.2 3.95 ± 0.31 98.7 ± 0.3 6.00 5.97 ± 0.11 99.5 ± 0.1 5.95 ± 0.25 99.2 ± 0.2 5.93 ± 0.15 98.8 ± 0.2 8.00 8.03 ± 0.16 100.4 ± 0.3 7.98 ± 0.19 99.8 ± 0.6 7.97 ± 0.16 99.6 ± 0.2 10.00 9.99 ± 0.37 99.9 ± 0.2 9.99 ± 0.15 99.9 ± 0.2 9.98 ± 0.24 99.8 ± 0.1 Mean ± SD = 99.4 ± 0.8 Mean ± SD = 99.8 ± 0.9 Mean ± SD = 99.2 ± 0.9