Development and Validation of an HPLC Method for Determination of Antidiabetic Drug Alogliptin Benzoate in Bulk and Tablets
Table 1
Results of method optimization.
| Column used | Mobile phase | Flow rate | Wavelength | Observation | Result |
| Restek C18, 125 × 4.0 mm i.d., 5 µm | (Buffer : methanol) (45 : 55) v/v | 1.0 ml/min | 216 nm | Poor resolution 1.4 | Method rejected | Thermo Scientific C18, 250 × 4.6 mm i.d., 5 µm | (Buffer : acetonitrile) (25 : 75) v/v | 1.0 ml/min | 277 nm | Poor resolution 1.6 | Method rejected | Thermo Scientific C18, 250 × 4.6 mm i.d., 5 µm | (Buffer : acetonitrile) (45 : 55) v/v | 1.0 ml/min | 277 nm | Good resolution 2.4 | Method accepted |
|
|