Research Article

Comparison of Three Sample Preparation Procedures for the Quantification of L-Arginine, Asymmetric Dimethylarginine, and Symmetric Dimethylarginine in Human Plasma Using HPLC-FLD

Table 2

Validation data: calibration curve equation, correlation coefficient, LLOQ, ULOQ, intra- and interday precision, and average concentration ± SD quality control samples (human plasma from 6 volunteers) for L-arginine and ADMA using the selected sample preparation (A) with the developed chromatographic method.

Calibration curve equationCorrelation coefficient (R2)LLOQ (µM)ULOQ (µM)Intraday precision (CV, %) (n = 6)Interday precision (CV, %) (n = 9)Quality control (µM) (n = 6)

L-argininey = 78686x − 35190.9990.14153.25.064.1 ± 10.3
ADMAy = 73655x + 11340.9940.0120.503.43.30.27 ± 0.02
SDMAy = 87003x + 32610.9894.23.80.57 ± 0.09