Research Article
Comparative Study on Two Pretreatment Processes for Chemical Phase Analysis of Gold in Geological Samples by Atomic Absorption Spectrometry
Table 1
Instrument operating parameters.
| Parameter | Setting |
| XRF | Initialization element | Ag | Initialization channel | 2210 | Tube current | 250 μA | Tube voltage | 40 kV | Counting rate | 1 | Vacuum time | 25 s | Measure time | 100 s | Replicates | 3 |
| Flame AAS | Element | Au | Wavelength | 242.8 nm | Spectral bandwidth | 0.4 nm | Lamp current | 2.0 mA | Filter coefficient | 0.6 | Integration time | 3 s | Burner height | 6 mm | Flame type | Air-C2H2 |
| Graphite furnace AAS | Element | Au | Wavelength | 242.8 nm | Spectral bandwidth | 0.4 nm | Lamp current | 4.0 mA | Filter coefficient | 0.1 | Integration time | 3 s | Injection volume | 20 μL | Measurement methods | Peak area |
|
|
XRF: X-ray fluorescence spectrometry; AAS: atomic absorption spectrometry.
|