Potentiometric Stripping Analysis of Cadmium and Lead with Constant Inverse Current in the Analytical Step Using an Open Tubular Mercury-Coated Glassy Carbon Electrode

Table 4

The analytical characteristics of the proposed methods.

ParameterValue
CdPb

Limit of quantitation (μg/L)0.110.82
Repeatability (%)2.532.71
Reproducibility (%)3.924.19
Linearity
 Range (μg/L)2–3040–100
 Slope (s·L/μg)0.0580.037
 Standard deviation of the slope (s·L/μg)0.0020.002
 Intercept (s)0.070.09
 Standard deviation of the intercept (s)0.0030.002
 Correlation coefficient0.9960.991