Research Article

A Sensitive Fluorescence Biosensor for Silver Ions (Ag+) Detection Based on C-Ag+-C Structure and Exonuclease III-Assisted Dual-Recycling Amplification

Table 2

Determination of Ag+ in Huguang lake water samples ().

SamplesAdded (pmol/)Proposed method (mean ± SD) (pmol/L)AAS (mean ± SD) (pmol/L)

15.05.1 ± 0.34.8 ± 0.1
250.052.9 ± 1.849.1 ± 1.4
3500.0506.7 ± 12.0503.4 ± 16.3
41500.01493.4 ± 28.51520.1 ± 34.5

Mean represents the average of three determinations; SD, standard deviation.