Research Article

A Sensitive Fluorescence Biosensor for Silver Ions (Ag+) Detection Based on C-Ag+-C Structure and Exonuclease III-Assisted Dual-Recycling Amplification

Table 3

Determination of Ag+ in human sera samples ().

SamplesAdded (pmol/)Proposed method (mean ± SD) (pmol/L)AAS (mean ± SD) (pmol/L)

15.05.4 ± 0.25.1 ± 0.1
250.048.3 ± 1.649.3 ± 1.7
3500.0511.2 ± 13.4506.9 ± 11.0
41500.01481.2 ± 26.81543.6 ± 21.3

Mean represents the average of three determinations; SD, standard deviation.