Speciation Analysis of Arsenic Compounds by High-Performance Liquid Chromatography in Combination with Inductively Coupled Plasma Dynamic Reaction Cell Quadrupole Mass Spectrometry: Application for Vietnamese Rice Samples

Table 1

Operating conditions of HPLC-ICP-DRC-QMS for speciation of arsenic compounds.

Operating conditions of HPLC-ICP-DRC-QMS

High-performance liquid chromatography operating conditions
PumpAltus A-30 quaternary pump equipped with solenoid solvent selection valve, online degasser, liquid autosampler with 100 µL injection loop, and column temperature compartment
ColumnPRP X 100, 4.6 × 150 mm, 5 μm particle size; Hamilton, NV, USA
Column temperatureRoom temperature (app. 25°C)
EluentChannel A: 5 mM (NH4)2CO3, pH 9.0, 0, 05% EDTA, 5% MeOH
Channel B: 50 mM (NH4)2CO3, pH 9.0, 0, 05% EDTA, 5% MeOH
Elution modeGradient elution: 0-1 minute: 100% A and then increasing linearly to 100% B in 10 minutes and kept for 4 minutes. Finally decreasing to 100% in 0.1 minutes and kept for 6 minutes to regenerate column
Flow rate1.0 mL·min−1
Injection volume50 μL
ICP-DRC-QMS operating conditions
RF power1600 W
Plasma gas flow15 L·min−1
Auxiliary gas flow1.2 L·min−1
Nebulizer gas1 L·min−1
NebulizerPFA
Spray chamberQuartz cyclonic
Sample conePlatinum sample cone
Skimmer conePlatinum hyper skimmer cone
Cell entrance voltage−7 V
Cell exit voltage−7 V
DRC modeDynamic reaction cell with oxygen gas, 0.7 mL·min−1
Monitored ionAsO+ (m/z 91)
Measurement modePeak hopping
Dwell time100 ms
RPq0.6
Quantification modeExternal calibration curve