Research Article

Functionalized Fe3O4 Magnetic Nanoparticle Potentiometric Detection Strategy versus Classical Potentiometric Strategy for Determination of Chlorpheniramine Maleate and Pseudoephedrine HCl

Table 5

Determination of CPM and PSE in pharmaceutical formulation by the proposed sensors and comparison with the reported method.

Pharmaceutical formulationSensor 3Sensor 6Reported method [23]b

Sinlerg® tablets labelled to contain 2 mg CPM, 30 mg PSE, and 200 mg IBF/tablet (batch no. 212314)Mean ± SDa102.26 ± 1.204101.09 ± 1.106100.63 ± 1.317
Student’s t test0.518 (2.26b)0.743 (2.26b)
F test1.623 (6.26b)1.182 (6.26b)

aAverage of five determinations. bReported HPLC method using C18 column; flow rate 1.5 mL·min−1; mobile phase is a gradient elution of acetonitrile:buffer (15 : 85, v/v) for 5.5 min, (45 : 55, v/v) for 5.5–12 min, and (60 : 40, v/v) for 12–17 min at pH = 3 and UV detection at 220 nm.