Research Article

Influence of Different Chromatographic Conditions on the Level of Detection and Quantitation of Spironolactone by means of TLC-Densitometry

Table 3

Limits of detection and quantitation of spironolactone obtained by means of the NP-TLC method in different chromatographic conditions.

NP-TLC method
SorbentMobile phaseLOD1LOD2Average value of LODLOQ1LOQ2Average value of LOQ
µg/spotµg/spot

1.05715A0.0520.1150.0840.1590.3500.254
B0.0430.0950.0690.1310.2870.209
C0.0630.3130.1880.1900.9480.569

1.05567A0.0180.0510.0340.0530.1530.103
B0.0440.1890.1160.1340.5730.354
C0.5140.7850.6501.5582.3801.969

1.05554A0.0550.2740.1640.1660.8290.498
B0.0410.0900.0660.1240.2720.198
C0.5050.7720.6381.5312.3381.934

Mobile phase A: n-hexane-ethyl acetate-glacial acetic acid (24.5 : 24.5 : 1, v/v/v); mobile phase B: chloroform-acetone (45 : 5, v/v); mobile phase C: ethyl acetate-n-hexane (38 : 12, v/v); 1the value calculated on the basis of residual standard deviation of the appropriate calibration curve; 2the value calculated on the basis of standard deviation of the intercept of the appropriate calibration curve.