Determination of Isosorbide-5-Mononitrate in Human Plasma by High-Performance Liquid Chromatography-Tandem Mass Spectrometry and Its Application to a Bioequivalence Study

Table 2

Intraday and interday accuracy and precision of quality control samples of 5-ISMN.

Nominal con. (ng/ml)Intraday (n = 6)Interday (n = 18)
Measured con. (mean ± SD, ng/mL)Precision (RSD %)Accuracy (%)Measured con. (mean ± SD, ng/mL)Precision (RSD %)Accuracy (%)a

54.64 ± 0.234.8-7.24.56 ± 0.306.6−8.8
88.22 ± 0.445.32.88.13 ± 0.445.41.6
8085.7 ± 4.014.77.182.2 ± 3.314.02.8
800824 ± 31.23.83.0817 ± 19.22.42.1

Concentration is in 3 significant figures; SD, standard deviation; RSD%, percent relative standard deviation (in one decimal place). aExpressed as ((mean measured concentration−nominal concentration)/(nominal concentration)) × 100 (in one decimal place).