Research Article

Determination of Isosorbide-5-Mononitrate in Human Plasma by High-Performance Liquid Chromatography-Tandem Mass Spectrometry and Its Application to a Bioequivalence Study

Table 4

Stability assessments of 5-ISMN in plasma (n = 6).

MatrixStorage conditionsNominal con. (ng/mL)Measured con. (mean ± SD, ng/mL)Precision (RSD %)Accuracy (%)a

BloodRoom temperature for 2 h87.76 ± 0.182.3−2.5
200212 ± 5.182.46.0

PlasmaThree circles of freeze/thaw88.60 ± 0.354.17.5
800833.2 ± 8.821.14.1
In autosampler for 72 h after extraction88.62 ± 0.435.17.8
800823.2 ± 24.533.02.9
Room temperature for 8 h87.90 ± 0.354.5−1.3
800823.5 ± 7.230.92.9
Long-term stability (−30°C for 63 days)87.49 ± 0.537.1−6.4
800820.5 ± 26.853.32.6
Long-term stability (−80°C for 91 days)88.31 ± 0.192.24.0
800755.67 ± 9.931.3−5.5
Repeat injection for 64 h after extraction87.71 ± 0.222.9−3.6
800821.7 ± 25.123.12.7

Concentration is in 3 significant figures; SD, standard deviation; RSD%, percent relative standard deviation (in one decimal place). aExpressed as ((mean measured concentration−nominal concentration)/(nominal concentration)) × 100 (in one decimal place).