Determination of Isosorbide-5-Mononitrate in Human Plasma by High-Performance Liquid Chromatography-Tandem Mass Spectrometry and Its Application to a Bioequivalence Study

Table 5

Precision and accuracy of 5-ISMN in hyperlipidemia and hemolysis matrices (n = 6).

MatrixNominal con. (ng/mL)Measured con. (mean ± SD, ng/mL)Precision (RSD %)Accuracy (%)a

300 mg/dL hyperlipidemic matrix87.79 ± 0.374.7−2.6
8082.5 ± 3.894.73.1
800831 ± 23.182.83.8

2% hemolysis matrix87.91 ± 0.465.8−1.1
8080.8 ± 2.743.41.0
800852 ± 14.151.76.5

Concentration is in 3 significant figures; SD, standard deviation; RSD%, percent relative standard deviation. aExpressed as ((mean measured concentration−nominal concentration)/(nominal concentration)) × 100 (in one decimal place).