| GC/MS | 7890A/5975C | Intuvo 9000/5977B | EM 640 |
| Column | HP-5MS 30 m × 0.25 mm, 0.25 μm | HP-5MS 30 m × 0.25 mm, 0.25 μm | HP-5MS 25 m × 0.35 mm, 1 μm | Carrier gas | Helium, 147 kPa constant pressure | Helium, 1.2 mL/min constant flow | Nitrogen, 500 hPa constant pressure | Sampler | Agilent GC 80 | Agilent 7693A | — | Injection volume | 1 μl | 1 μl | 1 μl | Inlet | 290°C, splitless mode, purge flow 100 mL/min at 2 min | 290°C, splitless mode, purge flow 100 mL/min at 2 min | 230°C, splitless mode, purge flow 30 mL/min at 1 min | Oven | 40°C (2 min), 10°C/min to 280°C (10 min) | 40°C (2 min), 10°C/min to 280°C (10 min) | 40°C (2 min), 10°C/min to 280°C (10 min) | Detector | Quadrupole MS, EI, scan mode, transfer line 290°C, scan range 35–600 amu, solvent delay 6 min | Quadrupole MS, EI, scan mode, transfer line 290°C, scan range 35–600 amu, solvent delay 6 min | Quadrupole MS, EI, scan mode, transfer line 280°C, scan range 50–550 amu, solvent delay 6 min | Evaluation of chromatograms | | Agilent chemstation GC/MSD—data analysis, version E.02.02., Agilent Technologies, Inc., 2011 | MassHunter Workstation software, version B.07.00, Agilent Technologies, Inc., 2014 | Bruker Data Analysis, version 1.1., Bruker Daltonik GmbH, 2003 |
|
|