Research Article

A Chemiluminescent Immunoassay for Osteocalcin in Human Serum and a Solution to the “Hook Effect”

Table 3

Precision test results.

SamplesIntra-assay CV (n = 20)Interassay CV (n = 20)
Mean (ng/mL)SDCV (%)Mean (ng/mL)SDCV (%)

119.981.095.4619.321.47.25
245.532.034.4644.832.645.89
380.872.63.2282.133.634.42