Research Article

High-Accuracy Determination of Potassium and Selenium in Human Serum by Two-Step Isotope Dilution ICPMS

Table 1

Instrument parameters.

ParameterKSe

ICP RF power (W)11001100
Gas flows (L/min)0.970.97
Lens voltage (V)6.756.75
Analog stage voltage (V)−1700−1700
Pulse stage voltage (V)900900
Dwell time per AMU2ms2ms
Scan modePeak hoppingPeak hopping
Sweeps/reading2525
Readings/replicate2025
Replicates1525
Cell gas A : CH4 (mL/min)1.50.8
RP q: Ar (mL/min)0.80.45