High-Accuracy Determination of Potassium and Selenium in Human Serum by Two-Step Isotope Dilution ICPMS

Table 3

Analysis of standard reference material by two-way ID-ICPMS.

K (mg kg−1)Se (μg kg−1)
NIST956DGBW09152NIST909CGBW09152
Level 1Level 2Level 3

Test 1218.43140.7862.52166.7116.376.3
Test 2219.66141.2162.39168.0115.976.5
Test 3219.74141.6961.04167.6115.875.2
Avg219.28141.2361.98167.43116.0076.00
Certified values219.87 ± 1.86142.47 ± 1.3761.50 ± 0.53168.3 ± 3.4115.9 ± 3.275.7 ± 1.6
Bias (%)0.3%0.9%−0.8%0.5%−0.1%−0.4%