Research Article

High-Accuracy Determination of Potassium and Selenium in Human Serum by Two-Step Isotope Dilution ICPMS

Table 5

Sources of uncertainty in the determination of K.

Sources of uncertaintyK
Level 1Level 2Level 3
Value (xi)uc (xi)Value (xi)uc (xi)Value (xi)uc (xi)

Type A uncertainties
RY0.0410.000170.0410.000170.0410.00017
Rz110.063110.063110.063
RBc0.970.00141.00.00151.10.0018
RB0.930.00131.00.00150.910.0014
CB/mg kg−10.140.160.160.160.210.16
K

Type B uncertainties
Cz/mg kg−1460.10460.10460.10
mzc/mg3.3 × 1030.0403.3 × 1030.0403.3 × 1030.040
mYc/mg3.4 × 1030.0403.4 × 1030.0403.4 × 1030.040
mY/mg3.4 × 1030.0403.4 × 1030.0403.4 × 1030.040
mx/mg2.1 × 1030.0401.1 × 1030.0408.0 × 1020.040
Combined type A and type B0.220.340.37
Degrees of freedom (Veff)111111
Coverage factor (k)222
Measured value/mg kg−1136.8178.0252.3
Expanded uncertainty (U ())0.140.210.23