Research Article

Selective Consecutive Determination of Desloratadine and Montelukast Sodium in Their Pure and Binary Dosage Form Based on Pencil Graphite Electrochemical Sensors

Table 1

The performance characteristics of the proposed described sensors.

ParameterDES.RNK graphite sensorMON.Cd. graphite sensorCombined sensor
DESMON

Slope (mV. decade−1)−30.1127.70−29.1629.79
Intercept345.93−10.17337.60−119.73
Linearity range (M)5.00 × 10−5 − 1.00 × 10−21.00 × 10−5 − 1.00 × 10−21.00 × 10−5 − 1.00 × 10−25.00 × 10−6 − 1.00 × 10−2
Correlation coefficient0.99960.99970.99980.9992
LODa (µM)0.0360.0180.0250.025
LOQ (µM)0.1090.0550.0750.076
Working pH range2.5–56–102.5–56–10
Response time (seconds)13172021
Lifetime (days)63494949
Accuracyb (R%)100.9599.6098.48101.27
Repeatabilityc (RSD%)1.530.310.980.44
Intermediate precisiond (RSD%)1.900.861.051.19

a.Lod 3.3 SD of intercept/slope, and LOQ = 10 ∗ SD/slope. bAverage of three determinations. cRepeatability: the intraday precision (n = 3 × 3) and average of three concentrations (5 ∗ 10−5, 5 ∗ 10−4, and 5 ∗ 10−3 M) were repeated three times within the day. dIntermediate precision: the interday precision (n = 3 × 3) and average of three concentrations (5 ∗ 10−5, 5 ∗ 10−4, and 5 ∗ 10−3 M) were repeated three times on two consecutive days.