Development of an Ozone-Assisted Sample Preparation Method for the Determination of Cu and Zn in Rice Samples

Table 1

Figures of merit for the OAE and MAE methods.

ParameterCuZn
OAEMAEOAEMAE

Linear range (µg L−1)Up to 5.0 mg L−1Up to 1.0 mg L−1
LOD (mg kg−1, n = 6)0.200.180.080.10
LOQ (mg kg−1, n = 6)0.650.600.260.33
Precision (RSD%, OAE n = 18 and MAE n = 10)3.03.12.93.3
Certified value (mg kg−1) (NIST 1568a)2.4 ± 0.319.4 ± 0.5
Obtained value (mg kg−1)2.41 ± 0.092.32 ± 0.0618.6 ± 0.619.8 ± 0.7
Trueness (R%, OAE n = 18 and MAE n = 10)100.496.795.9102.1

Results are expressed as mean ± standard deviation. The replicate number arises from performing the OAE six times (3 simultaneous sample treatments each time plus a blank) considering the device design. Recovery (%) = (obtained value/certified value) × 100.