Operating conditions for target compounds in MS/MS mode in Waters Xevo-TQD mass spectrometer.

No.CompoundsRT (min)Ratio
TQ/TC
Precursor ionProduct ionsCone (V)Collision energy (eV)

1PARA2.878.6151.992.03022
109.918
2PARA-D42.87156.01143015
3TRIM3.131.6291.0229.92824
260.924
4OFLO3.142.5362.3261.32520
318.322
5OFLO-D33.14365.0321.15520
6CIPRO3.1511.6332.1288.14218
314.122
7SULFA3.521.5253.992.03426
156.916
8SULFA–13C63.52260.098.13828
9CARBA3.811.2236.9178.94234
193.636
10KETO4.11254.9209.13612