Research Article

Quantitative Measurement of Plasma Free Metanephrines by a Simple and Cost-Effective Microextraction Packed Sorbent with Porous Graphitic Carbon and Liquid Chromatography-Tandem Mass Spectrometry

Table 1

Intra- and interday assay imprecision and accuracy (spiked QC samples: LQC, MQC, & HQC and Chromsystems QC: C1 & C2).

CompundSampleAdded/expected (pg/mL)IntradayInterday
Mean (ng/mL)SDCV (%)Accuracy (%)Mean (ng/mL)SDCV (%)Accuracy (%)

MNLQC100108109.4108.0108109.7108.0
MQC500481439.096.2516428.2103.2
HQC200018991507.995.020301527.5101.5
C16058.98.013.698.262.98.513.4104.8
C2750708628.794.4729567.797.2

NMNLQC1501411812.894.01511610.7100.7
MQC750811799.8108.1781688.7104.1
HQC350033101053.294.631001825.988.6
C110010388.0103.010998.6109.0
C235003080682.288.034602246.598.9