Development and Validation of a Method for the Analysis of Zinc Oxide in Cosmetic Matrices by Flame Atomic Absorption Spectroscopy

Table 4

Average found concentration, standard deviation, and total variation (%RSD) for the precision levels considered in this study.

Raw materialLipstickEmulsionFoundation

Repeatability (mean ± SD% w/w; %RSD)Level 13.155 ± 0.053; 1.696.629 ± 0.130; 1.955.593 ± 0.120; 2.15
Level 259.13 ± 0.69; 1.16a3.954 ± 0.066; 1.668.796 ± 0.113; 1.287.793 ± 0.091; 1.16
Level 34.627 ± 0.076; 1.649.784 ± 0.133; 1.369.613 ± 0.136; 1.42

Global intermediate precision (mean ± SD% w/w; %RSD)Day 156.16 ± 0.73; 1.303.628 ± 0.067; 1.858.310 ± 0.120; 1.458.469 ± 0.290; 3.43
Day 258.09 ± 0.77; 1.323.810 ± 0.036; 0.958.443 ± 0.071; 0.848.577 ± 0.146; 1.70
Global57.13 ± 1.23; 2.163.719 ± 0.108; 2.918.377 ± 0.117; 1.408.523 ± 0.226; 2.65

Global reproducibility (mean ± SD% w/w; %RSD)Laboratory 159.55 ± 0.33; 0.553.722 ± 0.055; 1.473.517 ± 0.022; 0.648.697 ± 0.051; 0.58
Laboratory 256.80 ± 1.36; 2.394.025 ± 0.166; 4.123.673 ± 0.089; 2.428.983 ± 0.095; 1.05
Global58.17 ± 1.75; 3.013.873 ± 0.200; 5.153.595 ± 0.103; 2.888.840 ± 0.171; 1.93

aRaw material was evaluated at its nominal concentration, as declared by the certificate of analysis.