Development and Validation of Near-Infrared Methods for the Quantitation of Caffeine, Epigallocatechin-3-gallate, and Moisture in Green Tea Production

Table 5

Validation results of the method.

ProjectDescribeCaffeineMoisture contentEGCG

LinearityNIR prediction = a reference value + bTraining setTraining setTraining set
Y = 0.9957X + 0.0223Y = 0.9812X − 1.7039Y = 0.9885X + 0.0944
R2 = 0.9957R2 = 0.9812R2 = 0.9885

AccuracyThe paired t-test between the NIR predicted value and the reference value of the sample in the validation setPrediction setPrediction setPrediction set
t = −1.441t = −0.663t = 0.505
p = 0.162 (α = 0.05)p = 0.513 (α = 0.05)p = 0.618 (α = 0.05)

RepeatabilityThe same sample was measured 6 times on the same day by the same operatorSample reference 3.45%Sample reference 6.05%Sample reference 7.15%
SD = 0.095SD = 0.202SD = 0.107
RSD = 2.699%RSD = 3.654%RSD = 1.459%

Intermediate precisionThe same sample was measured by two operators for three consecutive daysSample reference 3.45%Sample reference 6.05%Sample reference 7.15%
SD = 0.029SD = 0.179SD = 0.039
RSD = 0.835%RSD = 2.968%RSD = 0.554%

SD: standard deviation; RSD: relative standard deviation.