Research Article

Flame Atomic Absorption Determination of Gold Ion in Aqueous Samples after Preconcentration Using 9-Acridinylamine Functionalized γ-Alumina Nanoparticles

Table 2

Data of real sample analysis for Au on 9-acridinylamine alumina.

Real sample (ng mL−1)Added (ng mL−1)Found (ng mL−1)Recovery (%)

Distilled waterND5.004.95 ± 0.1099
Tap waterND5.004.89 ± 0.1098
Waste water12.31 40.0051.81 ± 0.1099
Sea water8.0040.0045.64 ± 0.1096
Standard material260.000.00265.33 ± 2.41102